Performance and patients' satisfaction with the A7+TouchCare insulin patch pump system: A randomized controlled non-inferiority study

PLOS ONE(2023)

引用 0|浏览7
暂无评分
摘要
BackgroundWe assessed the performance and patient satisfaction of a new insulin patch pump, the A7+TouchCare (Medtrum), compared with the Omnipod system.MethodsThis multicenter, randomized, open-label, controlled study enrolled 100 adult patients with type 1 or type 2 diabetes mellitus (A1C & GE; 6.5% and & LE; 9.5%, i.e., 48 to 80 mmol/mol) who were assigned with the Omnipod or with the A7+TouchCare pump for 3 months. The primary study outcome was the glucose management indicator (GMI) calculated with continuous glucose monitoring (CGM).ResultsPremature withdrawals occurs respectively in 2 and 9 participants in the Omnipod and TouchCare groups. In the Per Protocol analysis, the difference in GMI between groups was 0.002% (95% confidence interval -0.251; 0.255). The non-inferiority was demonstrated since the difference between treatments did not overlap the pre-defined non-inferiority margin (0.4%). There was no significant difference in CGM parameters between groups. On average, patients in both groups were satisfied/very satisfied with the insulin pump system. Patients preferred Omnipod as an insulin management system and especially the patch delivery system but preferred the A7+TouchCare personal diabetes manager to control the system.ConclusionsThis study showed that the A7+TouchCare insulin pump was as efficient as the Omnipod pump in terms of performance and satisfaction.Clinical trail registrationThe study was registered in the ClinicalTrials.gov protocol register (NCT04223973).
更多
查看译文
关键词
insulin,satisfaction,patients,non-inferiority
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要