Towards real-time digital pulse process algorithms for CsI(Tl) detector array at External Target Facility in HIRFL-CSR

Nuclear Science and Techniques(2023)

引用 0|浏览28
暂无评分
摘要
fully digital data acquisition system based on a field-programmable gate array (FPGA) was developed for a CsI(Tl) array at the External Target Facility (ETF)in the Heavy Ion Research Facility in Lanzhou(HIRFL). To process the CsI(Tl) signals generated by γ-rays and light-charged ions, a scheme for digital pulse processing algorithms is proposed. Every step in the algorithms was benchmarked using standard γ and α sources. The scheme, which included a moving average filter, baseline restoration, leading-edge discrimination, moving window deconvolution and digital charge comparison was subsequently implemented on the FPGA. A good energy resolution of 5.7% for 1.33 MeV γ rays and excellent α-γ identification using the digital charge comparison method were achieved, which satisfies CsI(Tl) array performance requirements.
更多
查看译文
关键词
CsI(Tl) array,on-line digital algorithms,moving average filter,moving window deconvolution,on-line particle identification algorithms