Phantom study of layered sensor module for photon-counting BMD detector

Physical and engineering sciences in medicine(2023)

引用 0|浏览10
暂无评分
摘要
In this study, we perform bone mineral density (BMD) calculation by designing a layered sensor module (LSM) that divides high- and low-energy spectra from a single shot of X-rays. Gamma-ray evaluation supports this mechanism; low-energy gamma rays are absorbed in the front detector, whereas high-energy gamma rays are absorbed in the rear detector. In this phantom study, LSM divides a single shot of X-ray into two spectra with different distributions of energy, thereby affording X-ray images with different properties, such as contrast and gray scale. The region of interest (ROI) is classified by the Prewitt operator to sort the pixels for BMD calculation or Rs value. The calculated final value is 1.2051 g/cm 2 with a standard deviation (SD) of 0.3690 g/cm 2 , as obtained from our previous study. An improved SD results from the layered structure with two channels for signal processing, the introduction of Rs value, and the use of Prewitt filter to sort reliable data. Overall, this study displays the feasibility of LSM for BMD calculation with a small error, thereby enabling the diagnosis of osteoporosis with novel mechanism.
更多
查看译文
关键词
Layered sensor module,Bone mineral density,CdTe-based detector