Validation study of new left atrial appendage closure device measurement by two-dimensional and three-dimensional transesophageal echocardiography

Journal of Echocardiography(2024)

引用 0|浏览5
暂无评分
摘要
Background In left atrial appendage closure using WATCHMAN FLX, accurate device measurement by transesophageal echocardiography (TEE) is important. We aimed to experimentally validate appropriate methods of device measurement with two-dimensional (2D) and three-dimensional (3D) TEE compared with actual size. Methods We prepared a full range of device sizes (20, 24, 27, 31, 35 mm), each with five different compression rates. Each device was measured by 2D and 3D TEE at depths of 2, 4, and 6 cm in vitro using inner, outer, and middle line methods. We compared the difference between the actual size by caliper and measurements at each compression rate and depth by the three methods in 2D and 3D TEE. Results A total of 450 patterns of measurements were analyzed. The differences using the middle line method were much less than those using the inner and outer line methods in 2D and 3D TEE (2D TEE: 0.45 ± 0.36 vs. 2.55 ± 0.99 vs. 2.59 ± 0.72 mm, p < 0.01; 3D TEE: 0.34 ± 0.27 vs. 2.38 ± 0.69 vs. 1.86 ± 0.77 mm, p < 0.01). Moreover, the differences in measurements by 3D TEE were more accurate than those of 2D TEE in the inner (2.47 ± 1.86 vs. 1.86 ± 0.77 mm, p < 0.01) and middle (0.58 ± 0.37 vs. 0.34 ± 0.27 mm, p < 0.01) line methods. Conclusions Middle line method by 3D TEE is the most reliable approach for device measurement at left atrial appendage closure using WATCHMAN FLX device.
更多
查看译文
关键词
Atrial fibrillation,Left atrial appendage,Left atrial appendage closure,Two-dimensional transesophageal echocardiography,Three-dimensional transesophageal echocardiography
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要