Research of integrated shimming Halbach magnet for High strength, compact Benchtop NMR device.

Journal of magnetic resonance (San Diego, Calif. : 1997)(2023)

引用 0|浏览5
暂无评分
摘要
Due to production and assembly errors of magnets resulting in reduced magnetic field homogeneity, passive shimming (PS) is necessary when Halbach magnets are used in high-resolution Benchtop Nuclear Magnetic Resonance (BNMR) spectrometer. The conventional PS technique, which places independent PS devices inside the magnet aperture, is no longer applicable to small-aperture compact high-field-strength Halbach magnet studies. In this paper, based on spherical harmonic function expansion, we improve the magnetic field homogeneity by optimizing the moving step of the magnet moving arrays composed of Halbach magnets to generate the corresponding harmonic terms to compensate for the magnetic field. With this approach, the homogeneity of a 1 T Halbach magnet was improved from the original 3913 ppm to 8 ppm in a L10 mm × R2.5 mm of 0.64% copper sulfate doped water sample. This work explores the PS mechanism based on the movement of magnetic blocks, which can be applied in BNMR and other compact high-field strength high-homogeneity Halbach magnets application circumstances.
更多
查看译文
关键词
halbach magnet