Effects of hybrid closed-loop system on glycemic control and psychological aspects in persons with type 1 diabetes treated with sensor-augmented pump: A prospective single-center observational study

Tomoaki Akiyama,Tadashi Yamakawa,Kazuki Orime, Masahiro Ichikawa, Marina Harada, Takumi Netsu, Ryoichi Akamatsu, Keita Nakamura, Satoru Shinoda,Yasuo Terauchi

JOURNAL OF DIABETES INVESTIGATION(2024)

引用 0|浏览10
暂无评分
摘要
Aims/introduction: This study evaluated the effects of the Medtronic MiniMed 770G hybrid closed-loop system on glycemic control and psychological aspects in persons with type 1 diabetes mellitus.Materials and methods: This 3-month prospective observational study included 22 participants with type 1 diabetes mellitus who used the Medtronic MiniMed 640G predictive low-glucose suspend system and were switched to the 770G system. Time in the range of 70-180 mg/dL and glycated hemoglobin levels were evaluated; satisfaction, emotional distress and quality of life were assessed using self-reported questionnaires, including the Diabetes Treatment Satisfaction Questionnaire Status, Problem Area in Diabetes and Diabetes Therapy-Related Quality of Life.Results: Time in the range of 70-180 mg/dL increased (63.5 +/- 13.4 to 73.0 +/- 10.9% [mean +/- standard deviation], P = 0.0010), and time above the range of 181-250 mg/dL decreased (26.9 +/- 8.9 to 19.6 +/- 7.1%, P < 0.0005). Glycated hemoglobin levels decreased (7.7 +/- 1.0 to 7.2 +/- 0.8%, P = 0.0021). The percentage of participants with time below the range of 54-69 mg/dL <4% of readings increased from 91% to 100% (P < 0.0005). No significant changes were detected in the satisfaction, emotional distress and quality of life levels, but increased sensor calibration might be related to worsened emotional distress and quality of life.Conclusions: The hybrid closed-loop system decreased hyperglycemia and minimized hypoglycemia, but did not improve psychological aspects compared with the predictive low-glucose suspend system, probably because sensor calibration was increased.
更多
查看译文
关键词
Blood glucose,Insulin infusion systems,Type 1 diabetes mellitus
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要