A Predictive and an Optimization Mathematical Model for Device Design in Cardiac Pulsed Field Ablation Using Design of Experiments.

Journal of cardiovascular development and disease(2023)

引用 0|浏览5
暂无评分
摘要
Cardiac catheter ablation (CCA) is a common method used to correct cardiac arrhythmia. Pulsed Field Ablation (PFA) is a recently-adapted CCA technology whose ablation is dependent on electrode and waveform parameters (factors). In this work, the use of the Design of Experiments (DoE) methodology is investigated for the design and optimization of a PFA device. The effects of the four factors (input voltage, electrode spacing, electrode width, and on-time) and their interactions are analyzed. An empirical model is formed to predict and optimize the ablation size responses. Based on the ranges tested, the significant factors were the input voltage, the electrode spacing, and the on time, which is in line with the literature. Two-factor interactions were found to be significant and need to be considered in the model. The resulting empirical model was found to predict ablation sizes with less than 2.1% error in the measured area and was used for optimization. The findings and the strong predictive model developed highlight that the DoE approach can be used to help determine PFA device design, to optimize for certain ablation zone sizes, and to help inform device design to tackle specific cardiac arrhythmias.
更多
查看译文
关键词
pulsed field ablation, atrial fibrillation, in vitro, cardiomyocytes, design of experiments
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要