Chrome Extension
WeChat Mini Program
Use on ChatGLM

Simultaneous Monitoring of Amoxicillin and Paracetamol in Synthetic Biological Fluids Using a 3D Printed Disposable Electrode with a Lab-Made Conductive Filament

Analytical and bioanalytical chemistry/Analytical & bioanalytical chemistry(2023)

Cited 0|Views7
No score
Abstract
In this work, we are pleased to present for the first time a 3D-printed electrochemical device using a lab-made conductive filament based on graphite (Gr) and polylactic acid (PLA) polymer matrix for the simultaneous detection of amoxicillin (AMX) and paracetamol (PAR). The sensor was properly characterized by scanning electron microscopy (SEM), electrochemical impedance spectroscopy (EIS), and cyclic voltammetry (CV). Compared to the commercial glassy carbon electrode (GCE), the superior performance of the 3D-Gr/PLA electrode was verified with a 3.8-fold more favored charge transfer. A differential pulse voltammetry (DPV) method was proposed providing a linear working range of 4 to 12 μmol L-1 for both analytes and a limit of detection (LOD) of 0.80 and 0.51 μmol L-1 for AMX and PAR, respectively. Additionally, repeatability studies (n = 5, RSD < 5.7
More
Translated text
Key words
Lab-made conductive filament,Simultaneous monitoring,Amoxicillin analysis,Paracetamol analysis,3D-printed devices
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined