Comparison of two deep-learning image reconstruction algorithms on cardiac CT images: A phantom study

Joel Greffier, Maxime Pastor,Salim Si-Mohamed, Cynthia Goutain-Majorel, Aude Peudon-Balas, Mourad Zoubir Bensalah,Julien Frandon,Jean-Paul Beregi,Djamel Dabli

DIAGNOSTIC AND INTERVENTIONAL IMAGING(2024)

引用 0|浏览2
暂无评分
摘要
Purpose: The purpose of this study was to compare the performance of Precise IQ Engine (PIQE) and Advanced intelligent Clear-IQ Engine (AiCE) algorithms on image-quality according to the dose level in a cardiac computed tomography (CT) protocol. Materials and methods: Acquisitions were performed using the CT ACR 464 phantom at three dose levels (volume CT dose indexes: 7.1/5.2/3.1 mGy) using a prospective cardiac CT protocol. Raw data were reconstructed using the three levels of AiCE and PIQE (Mild, Standard and Strong). The noise power spectrum (NPS) and task-based transfer function (TTF) for bone and acrylic inserts were computed. The detectability index (d') was computed to model the detectability of the coronary lumen (350 Hounsfield units and 4-mm diameter) and non-calcified plaque (40 Hounsfield units and 2-mm diameter). Results: Noise magnitude values were lower with PIQE than with AiCE (-13.4 +/- 6.0 [standard deviation (SD)] % for Mild, -20.4 +/- 4.0 [SD] % for Standard and -32.6 +/- 2.6 [SD] % for Strong levels). The average NPS spatial frequencies shifted towards higher frequencies with PIQE than with AiCE (21.9 +/- 3.5 [SD] % for Mild, 20.1 +/- 3.0 [SD] % for Standard and 12.5 +/- 3.5 [SD] % for Strong levels). The TTF values at fifty percent (f50) values shifted towards higher frequencies with PIQE than with AiCE for acrylic inserts but, for bone inserts, f50 values were found to be close. Whatever the dose and DLR level, d' values of both simulated cardiac lesions were higher with PIQE than with AiCE. For the simulated coronary lumen, d' values were better by 35.1 +/- 9.3 (SD) % on average for all dose levels for Mild, 43.2 +/- 5.0 (SD) % for Standard, and 62.6 +/- 1.2 (SD) % for Strong levels. Conclusion: Compared to AiCE, PIQE reduced noise, improved spatial resolution, noise texture and detectability of simulated cardiac lesions. PIQE seems to have a greater potential for dose reduction in cardiac CT acquisition. (c) 2023 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
更多
查看译文
关键词
Cardiac CT,Deep-learning image reconstruction,algorithm,Multidetector computed tomography,Task-based image quality assessment
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要