Test-retest repeatability and agreement of the quantitative contrast sensitivity function test: towards the validation of a new clinical endpoint

Graefe's Archive for Clinical and Experimental Ophthalmology(2024)

引用 0|浏览5
暂无评分
摘要
Purpose The purpose of this study is to investigate test-retest reliability and agreement of the quantitative contrast sensitivity function test (qCSF) in the retina clinic. Methods A total of 121 right eyes of 121 patients were tested and consecutively re-tested with qCSF in the retina clinic. Outcomes included area under the logarithm of contrast sensitivity function curve (AULCSF), contrast acuity, and contrast sensitivity thresholds at 1–18 cycles per degree (cpd). Test-retest means were compared with paired t -test, variability was compared with the Brown-Forsythe test, and intraclass correlation coefficient (ICC) and Bland Altman plots evaluated reliability and agreement. Results Mean test-retest differences for all qCSF metrics ranged from 0.02 to 0.05 log units without statistically significant differences in variability. Standard deviations ranged from 0.08 to 0.14. Coefficients of repeatability ranged from 0.16 to 0.27 log units. ICC > 0.9 for all metrics except 1cpd (ICC = 0.84, all p < 0.001); AULCSF ICC = 0.971. Conclusion qCSF-measured contrast sensitivity shows great test-retest repeatability and agreement in the retina clinic.
更多
查看译文
关键词
Contrast sensitivity,Repeatability,Agreement,Validation,Retina,Quantitative contrast sensitivity function,Visual function,Functional vision,Clinical endpoints