Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study

RADIATION PROTECTION DOSIMETRY(2024)

引用 0|浏览2
暂无评分
摘要
To determine whether using lower-tube voltage reduces the scattered dose for the lens during paediatric thoracic computed tomography (CT). Two paediatric anthropomorphic phantoms (ATOM Phantom, CIRS, Norfolk, Virginia, USA) representing a newborn and 5-year-old were placed on the gantry of CT scanner, and optically stimulated luminescence dosemeters were placed on the left and right lenses, in front of the left and right thyroid glands, in front of the left and right mammary glands, and in front of and behind the mammary gland level and we measured scattered dose of the optically stimulated luminescence dosemeter was compared for each phantom between 80 and 120 kVp. Significant differences were observed in the scatter doses for the lens between 80 and 120 kVp (p < 0.01). Compared with the 120 kVp scan, the scatter doses for the lens were similar to 15-40% lower in newborn and 5-year-olds using the 80 kVp scan during paediatric CT.