Validation of the Wellvii VitalDetect blood pressure monitor in general population according to the International Standardization Organization 81060-2:2018

Lin-Yi Li, Jin-Feng Chen,Sen Bing,Kui Liu, Rui Cheng,Yi Wan

BLOOD PRESSURE MONITORING(2024)

引用 0|浏览3
暂无评分
摘要
Objective: To evaluate the accuracy of the Wellvii VitalDetect automated oscillometric finger blood pressure monitor (single cuff size) for self/home blood pressure measurement according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018). Methods: According to the universal standard, a total of 92 participants were recruited and finally blood pressure of 85 eligible participants was sequentially measured and compared with a standard mercury sphygmomanometer. Results: A total of 255 comparison pairs were obtained and analyzed based on the universal standard. For the validation criterion 1 of the ISO 81060-2:2018 universal standard, the mean +/- SD of the differences between the test device and reference blood pressure readings was 1.66 +/- 7.67 and 1.04 +/- 6.45 mmHg for systolic and diastolic blood pressure, respectively. For criterion 2, the SD of the averaged blood pressure differences between the test device and reference blood pressure per subject was +/- 6.49 mmHg (pass <= 6.73 mmHg) and +/- 5.67 mmHg (pass <= 6.86 mmHg) for systolic and diastolic blood pressure, respectively. Conclusion: The Wellvii VitalDetect automated finger blood pressure monitor passed all the requirements for validation by the ISO 81060-2:2018 universal standard and can be recommended for self/home blood pressure measurement in general population. Copyright (c) 2023 Wolters Kluwer Health, Inc. All rights reserved.
更多
查看译文
关键词
blood pressure measurement,International Standardization Organization,ISO 81060-2:2018,universal standard,validation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要