Ultra-fast whole-body bone tomoscintigraphies achieved with a high-sensitivity 360° CZT camera and a dedicated deep-learning noise reduction algorithm

Achraf Bahloul,Antoine Verger, Yechiel Lamash, Nathaniel Roth, Diawad Dari,Pierre-Yves Marie,Laetitia Imbert

European Journal of Nuclear Medicine and Molecular Imaging(2023)

引用 0|浏览0
暂无评分
摘要
This study aimed to determine whether the whole-body bone Single Photon Emission Computed Tomography (SPECT) recording times of around 10 min, routinely provided by a high-sensitivity 360° cadmium and zinc telluride (CZT) camera, can be further reduced by a deep-learning noise reduction (DLNR) algorithm. Methods DLNR was applied on whole-body images recorded after the injection of 545 ± 33 MBq of [ 99m Tc]Tc-HDP in 19 patients (14 with bone metastasis) and reconstructed with 100%, 90%, 80%, 70%, 60%, 50%, 40%, and 30% of the original SPECT recording times. Results Irrespective of recording time, DLNR enhanced the contrast-to-noise ratios and slightly decreased the standardized uptake values of bone lesions. Except in one markedly obese patient, the quality of DLNR processed images remained good-to-excellent down to 60% of the recording time, corresponding to around 6 min SPECT-recording. Conclusion Ultra-fast SPECT recordings of 6 min can be achieved when DLNR is applied on whole-body bone 360° CZT-SPECT.
更多
查看译文
关键词
Bone scintigraphy,Whole-body CZT-camera,Deep learning,Recording time