High sensitivity air-coupled MHz frequency ultrasound detection using on-chip microcavities

arXiv (Cornell University)(2022)

引用 0|浏览1
暂无评分
摘要
Owing to their dual-resonance enhanced sensitivity, cavity optomechanical systems provide an ideal platform for ultrasound sensing. In this work, we realize high sensitivity air-coupled ultrasound sensing from kilohertz (kHz) to megahertz (MHz) frequency range based on whispering gallery mode microcavities. Using a 57 um-diameter microtoroid with high optical Q factor (~10^7) and mechanical Q factor (~700), we achieve sensitivities of 46 uPa Hz^{-1/2}-10 mPa Hz^{-1/2} in a frequency range of 0.25-3.2 MHz. Thermal-noise-limited sensitivity is realized around the mechanical resonance at 2.56 MHz, in a frequency range of 0.6 MHz. We also observe the second- and third-order mechanical sidebands, and quantitatively study the intensities of each mechanical sideband as a function of the mechanical displacement. Measuring the combination of signal to noise ratios at all sidebands has the potential to extend the dynamic range of ultrasound sensing. In addition, to improve the ultrasound sensitivity in the kHz frequency range, we use a microdisk with a diameter of 200 um, and achieve sensitivities of 1.83 uPa Hz^{-1/2}-10.4 mPa Hz^{-1/2} in 30 kHz-1.65 MHz range.
更多
查看译文
关键词
mhz frequency ultrasound detection,air-coupled,on-chip