Real-World Experience With the Angio-Seal Closure Device: Insights From Manufacturer and User Facility Device Experience Database

JOURNAL OF ENDOVASCULAR THERAPY(2023)

引用 0|浏览2
暂无评分
摘要
Purpose: Angio-Seal (Terumo Medical Corporations, Somerset, New Jersey) device is indicated for femoral arteriotomy closure. Real-world published data on complications are limited. We present 1 year of safety events involving Angio-Seal from the US Food and Drug Administration's post-market surveillance database of Manufacturer and User Facility Device Experience (MAUDE). Steps for managing frequent device-related problems are discussed. Materials and Methods: Angio-Seal MAUDE data from November 2019 to December 2020 was classified according to (1) mode of device failure, (2) complication, (3) treatment, and (4) Cardiovascular and Interventional Radiological Society of Europe (CIRSE) adverse event classification system. Results: There were 715 safety events, involving Angio-Seal VIP (93.1%), Evolution (5.7%), STS Plus (1.1%), and sizes 6F (62.5%) and 8F (37.5%). Failure mode involved unrecognized use of a damaged device (43.4%), failed deployment (20.1%), failed arterial advancement (6.3%), detachment of device component (4.9%), failed retraction (3.6%), operator error (1.1%), and indeterminate (20.6%). Of total, 44.8% of events were associated with patient harm. Complications involved minor blood loss (34.1%), hematoma (5.6%), significant blood loss (1.4%), and pseudoaneurysm (1.4%). Of total, 43.3% of cases required manual compression (MC), whereas 8.8% required more advanced intervention. Interventions included surgical repair (49.2%), thrombin injection (9.5%), balloon tamponade (6.3%), covered stent (4.8%), and unspecified (30.2%). Majority of safety events were CIRSE grade 1 (92.0%), followed by grades 2 (3.1%), 3 (4.6%), and 6 (deaths, 0.3%). Minority of devices were returned for manufacturer analysis (27.8%). Conclusions: The majority of safety events were associated with minor blood loss or local hematoma and could be addressed with MC alone. Most events were attributed to damaged device; however, very few devices were returned to manufacturer for analysis. This should be encouraged to allow for root cause analysis in order to improve safety profile of devices. System-level strategies for addressing barriers to under-reporting of safety events may also be considered.
更多
查看译文
关键词
Angio-Seal,MAUDE,vascular closure devices,safety,adverse event
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要