Decreasing implant load indicates spinal fusion when measured continuously

JOURNAL OF BIOMECHANICS(2024)

引用 0|浏览2
暂无评分
摘要
Reliable and timely assessment of bone union between vertebrae is considered a key challenge after spinal fusion surgery. Recently, a novel sensor concept demonstrated the ability to objectively assess posterolateral fusion based on continuous implant load monitoring. The aim of this study was to investigate systematically the concept in a mono-segmental fusion model using an updated sensor setup. Three sheep underwent bilateral facetectomy at level L2-L3 and L4-L5. The segments were stabilized using two unconnected pedicle-screw-rod constructs per level. Sensing devices were attached to the rods between each pedicle screw pair and the loads were continuously monitored over 16 weeks. After euthanasia, the spines were biomechanically tested for their range of motion and high-resolution CT scans were performed to confirm the fusion success. After an initial increase in implant load until reaching a maximum (100 %) at approximately week 4, eleven out of twelve sensors measured a constant decrease in implant load to 52 +/- 9 % at euthanasia. One sensor measurement was compromised by newly forming bone growing against the sensor clamp. Bridging bone at each facet and minor remnant segmental motion (<0.7 degrees) confirmed the fusion of all motion segments. Data obtained by continuous measurement of implant loading of spinal screw-rod constructs enables objective monitoring of spinal fusion progression. The sensor concept provides valuable real-time information, offering quantifiable data as an alternative to traditional imaging techniques. However, the design of the current sensor concept needs to be matured, tailored to, and validated for the human spine.
更多
查看译文
关键词
Spinal fusion,Continuous measurement,Implant load,Fusion assessment,In vivo measurements
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要