Stabilization of Traumatic Iliosacral Instability Using Innovative Implants: A Biomechanical Comparison

JOURNAL OF CLINICAL MEDICINE(2024)

引用 0|浏览0
暂无评分
摘要
(1) Background: Demographic changes over the past decade have had a significant impact on pelvic ring fractures. They have increased dramatically in the orthogeriatric population. Surgeons are faced with implant fixation issues in the treatment of these fragility fractures. This study compares two innovative implants for stabilizing the iliosacral joint in a biomechanical setting. (2) Methods: An iliosacral screw with a preassembled plate allowing the placement of an additional short, angular stable screw in the ilium and a triangular fixation system consisting of a fenestrated ilium screw and an iliosacral screw quasi-statically inserted through the "fenestra" were instrumented in osteoporotic artificial bone models with a simulated Denis zone 1 fracture. Biomechanical testing was performed on a servo-hydraulic testing machine using increasing, synchronous axial and torsional sinusoidal cyclic loading to failure. (3) Results: The SI-Plate and TriFix showed comparable stiffness values. The values for fracture gap angle and screw tip cutout were significantly lower for the TriFix compared to the SI-Plate. In addition, the number of cycles to failure was significantly higher for the TriFix. (4) Conclusions: Implant anchorage and primary stability can be improved in iliosacral instability using the triangular stabilization system.
更多
查看译文
关键词
dorsal pelvic ring,biomechanic,SI-plate,triangular fixation,iliosacral instability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要