Upgrade of the small-angle neutron scattering diffractometer SANS-J at JRR-3

JOURNAL OF APPLIED CRYSTALLOGRAPHY(2023)

引用 0|浏览2
暂无评分
摘要
The small-angle neutron scattering diffractometer SANS-J at the research reactor JRR-3, Tokai, Japan has been upgraded. A He-3 front detector was installed adjacent to the direct beam just behind the sample position to cover scattering in the range 1.3-6 nm(-1) for the magnitude of the scattering vector Q. Additionally, the maximum Q covered by the existing main He-3 detector was extended to 1.8 nm(-1) by decreasing the minimum sample-to-detector distance. These modifications enabled gapless measurements of the scattering curve in the Q range 0.002-6 nm(-1) with three standard device layouts that can be changed within 10 min. A new graphical user interface simplifies the operation of the device layouts and data acquisition.
更多
查看译文
关键词
small-angle neutron scattering instruments,SANS-J upgrade,wide-Q-range measurements,user-friendly GUI systems