Atrial Fibrillation Detection from Compressed ECG Measurements for Wireless Body Sensor NetworkJust Accepted

ACM Transactions on Internet Technology(2023)

引用 0|浏览5
暂无评分
摘要
Recent years have witnessed an increasing prevalence of wearable devices in the public, where atrial fibrillation (AF) detection is a popular application in these devices. Generally, AF detection is performed on cloud whereas this paper describes an on-device AF detection method. Technically, compressed sensing (CS) is first used for electrocardiograph (ECG) acquisition. Then QRS detection is proposed to be performed directly on the compressed CS measurements, rather than on the reconstructed signals on the powerful cloud server. Based on the extracted QRS information, AF is determined by quantitatively analyzing the ( RR , dRR ) plot. Databases with ECG samples collected from both medical-level (MIT-BIH afdb) and wearable ECG devices (Physionet Challenge 2017) are introduced for performance validation. The experiment results well demonstrate that our on-device AF detection algorithm can approach the performance of those implemented on the raw signals. Our proposal is suitable for AF screening directly on the wearable devices, without the support of the data center for signal reconstruction and intelligent analysis.
更多
查看译文
关键词
Atrial fibrillation,on-device detection,compressed sensing,wireless body sensor network
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要