Towards liquid EPR dosimetry using nitroxides in aqueous solution

PHYSICS IN MEDICINE AND BIOLOGY(2024)

引用 0|浏览3
暂无评分
摘要
Objective. Water-equivalent dosimeters are desirable for dosimetry in radiotherapy. The present work investigates basic characteristics of novel aqueous detector materials and presents a signal loss approach for electron paramagnetic resonance (EPR) dosimetry. Approach. The proposed principle is based on the radiation dose dependent annihilation of EPR active nitroxides (NO center dot) in aqueous solutions. Stable nitroxide radicals (3-Maleimido-2,2,5,5-tetramethyl-1-pyrrolidinyloxy (MmP), 3-Carbamoyl-2,2,5,5-tetramethyl-1-pyrrolidinyloxy (CmP)) in aqueous solutions containing dimethyl sulfoxide (DMSO) as an additive were filled in glass capillaries for irradiation and EPR readout. Radiation doses ranging from 1 to 64 Gy were applied with a clinical 6 MV flattening filter free photon beam. EPR readout was then performed with a X-band benchtop spectrometer. The dose response, temporal stability and reproducibility of the samples' EPR signal amplitudes as well as the influence of the nitroxide concentration between 10 and 160 mu M on the absolute signal loss were investigated using MmP. CmP was used to examine the dependence of the dose response on DMSO concentration between 0 and 10 vol%. An indirect effect model was fitted to the experimental data assuming irradiation induced radical reactions as the underlying mechanism. Main results. For an initial MmP concentration of 20 mu M, absolute EPR signal loss is linear up to a dose of 16 Gy with a yield G(-NO center dot) of approximately 0.4 mu mol J-1. Within five weeks upon sample irradiation to doses between 0 and 32 Gy relative EPR signal fluctuations were on average (126 readouts) below 1% (1 sigma). For c(MmP) >= 20 mu M, absolute signal loss is only weakly dependent on c(MmP), whereas it increases strongly with increasing c(DMSO) in the range 0-5 vol%. An indirect effect model is applicable to describe the reaction mechanism resulting in the observed dose response curve. Significance. Liquids consisting of nitroxides in aqueous solution and small amounts of DMSO (2 vol%) show promising basic characteristics for application as water-equivalent EPR dosimeter materials in radiotherapy. The EPR signal loss is based on an indirect effect mediated by diffusing radicals originating from the radiolysis of the water/DMSO mixture.
更多
查看译文
关键词
EPR dosimetry,radiotherapy,water equivalence,nitroxide,dimethyl sulfoxide
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要