First (calibration) experiment using proton beam from FRENA at SINP
arxiv(2024)
摘要
This work presents the first calibration experiment of a 3 MV Tandetron
accelerator, FRENA, performed in May 2022. The ^7Li(p,n) reaction threshold
was measured to calibrate the terminal voltage measuring device. A LiF target
of thickness 175 μg/cm^2 was used in the experiment. The measured
threshold was 1872±2.7 keV, indicating 6-10 keV energy shift.