A Microflow Cytometer Enabled by Monolithic Integration of a Microreflector with an Acoustic Resonator

Yaping Wang,Wei Wei, Xieruiqi Guan,Yang Yang, Bingyi Tang,Wenlan Guo,Chen Sun,Xuexin Duan

ACS SENSORS(2024)

引用 0|浏览2
暂无评分
摘要
Current microflow cytometers suffer from complicated fluidic integration and low fluorescence collection efficiency, resulting in reduced portability and sensitivity. Herein, we demonstrated a new flow cell design based on an on-chip monolithically integrated microreflector with a bulk acoustic wave resonator (MBAW). It enables simultaneous 3D particle focusing and fluorescence enhancement without using shear flow. Benefited by the on-chip microreflector, the captured fluorescence intensity was 1.8-fold greater than that of the Si substrate and 8.3-fold greater than that of the SiO2 substrate, greatly improving the detection sensitivity. Combined with the contactless acoustic streaming-based focusing, particle sensing with a coefficient of variation as low as 6.1% was achieved. We also demonstrated the difference between live and dead cells and performed a cell cycle assay using the as-developed microflow cytometry. This monolithic integrated MBAW provides a new type of opto-acoustofluidic system and has the potential to be a highly integrated, highly sensitive flow cytometer for applications such as in vitro diagnostics and point of care.
更多
查看译文
关键词
opto-acoustofluidic,microflow cytometer,microreflector,on-chip,acoustic streaming
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要