Evaluating Image Quality and Radiation Dose in Low-Dose Thoraco-Abdominal CT Angiography with a Tin Filter for Patients with Aortic Disease

Chang Hoon Oh,Soo Buem Cho, Hyeyoung Kwon

JOURNAL OF CLINICAL MEDICINE(2024)

引用 0|浏览0
暂无评分
摘要
Background: We aimed to compared radiation exposure and image quality between tin-filter-based and standard dose thoraco-abdominal computed tomography angiography (TACTA) protocols, aiming to address a gap in the existing literature. Methods: In this retrospective study, ninety consecutive patients undergoing TACTA were included. Of these, 45 followed a routine standard-dose protocol (ST100kV), and 45 underwent a low-dose protocol with a tin filter (TF100kV). Radiation metrics were compared. The signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM) were calculated for the thoracic and abdominal aorta and right common iliac artery. Two independent readers assessed the image noise, image contrast, sharpness, and subjective image quality. Results: The mean dose for the TF100kV group was significantly lower (DLP 128.25 +/- 18.18 mGy*cm vs. 662.75 +/- 181.29, p < 0.001; CTDIvol 1.83 +/- 0.25 mGy vs. 9.28 +/- 2.17, p = 0.001), with an effective dose close to 2.3 mSv (2.31 +/- 0.33 mSv; p < 0.001). The TF100kV group demonstrated greater dose efficiency (FOM, thoracic aorta: 36.70 +/- 22.77 vs. 13.96 +/- 13.18 mSv(-1), p < 0.001) compared to the ST100kV group. Conclusions: Dedicated low-dose TACTA using a tin filter can significantly reduce the radiation dose while maintaining sufficient diagnostic image quality.
更多
查看译文
关键词
tin filter,thoraco-abdominal CT angiography,radiation exposure,image quality,diagnostic imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要