Beam quality and the mystery behind the lower percentage depth dose in grid radiation therapy

Scientific Reports(2024)

引用 0|浏览10
暂无评分
摘要
Grid therapy recently has been picking momentum due to favorable outcomes in bulky tumors. This is being termed as Spatially Fractionated Radiation Therapy (SFRT) and lattice therapy. SFRT can be performed with specially designed blocks made with brass or cerrobend with repeated holes or using multi-leaf collimators where dosimetry is uncertain. The dosimetric challenge in grid therapy is the mystery behind the lower percentage depth dose (PDD) in grid fields. The knowledge about the beam quality, indexed by TPR 20/10 (Tissue Phantom Ratio), is also necessary for absolute dosimetry of grid fields. Since the grid may change the quality of the primary photons, a new 𝐤_𝐪,𝐪_0 should be evaluated for absolute dosimetry of grid fields. A Monte Carlo (MC) approach is provided to resolving the dosimetric issues. Using 6 MV beam from a linear accelerator, MC simulation was performed using MCNPX code. Additionally, a commercial grid therapy device was used to simulate the grid fields. Beam parameters were validated with MC model for output factor, depth of maximum dose, PDDs, dose profiles, and TPR 20/10. The electron and photon spectra were also compared between open and grid fields. The d max is the same for open and grid fields. The PDD with grid is lower ( 10
更多
查看译文
关键词
Grid therapy,Small field dosimetry,Spectrometry,Monte Carlo,Linac