The importance of ion kinetic energy for interference removal in ICP-MS/MS.

Amanda D French, Kali M Melby, Kirby P Hobbs, Richard M Cox, Greg Eiden,Eric W Hoppe, Isaac J Arnquist,Khadouja Harouaka

Talanta(2024)

引用 0|浏览6
暂无评分
摘要
The effect of ion kinetic energy on gas phase ion reactivity with ICP-MS/MS was investigated in order to explore tuning strategies for interference removal. The collision/reaction gases CO2, N2O and O2 were used to observe the ion product distribution for 48 elements using an Agilent tandem ICP-MS (ICP-MS/MS) as a function of reaction gas flow rate (pressure) and ion kinetic energy. The kinetic energy of the incident ion was varied by adjusting the octopole bias (Voct). The three gases all form oxides (MO+) as the primary product with differing reaction enthalpies that result in distinct differences in the ion energies required for reaction with product ion distributions that vary with Voct. Consequently, by varying the ion kinetic energy (i.e., Voct), differences in interference reactivity can be used to achieve maximum separation. Three practical application examples were reported to demonstrate how the ion kinetic energy can be varied to achieve the ideal ion product distribution for interference resolution: CO2 for the removal of 238U in Pu analyses, CO2 for the removal of 40Ar16O vs. 56Fe, and O2 for the removal of Sm in Eu analyses, analogous to Pu/Am. The results demonstrate how the starting ion energy defined by Voct is an important factor to fully leverage the utility of any given reaction gas to remove interferences in the mass spectrum using ICP-MS/MS.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要