An 8-channel low power ASIC for Helium-3 tube position sensitive neutron detectors

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2024)

引用 0|浏览6
暂无评分
摘要
The majority of neutron spectrometers designated for use in the China Spallation Neutron Source (CSNS) rely on position -sensitive detectors that utilize Helium -3 tube detector technology. To minimize the impact of air on neutron scattering experiments, detectors must be located inside a vacuum chamber. For these types of spectrometers, it is also advisable to house the readout electronics within the vacuum chamber to limit the number of feed -through cables required and enhance the signal-to-noise ratio. This paper proposes an application -specific integrated circuit (ASIC) dedicated to position -sensitive Helium -3 tube detectors, including front-end amplification, shaping module and readout driving buffer. The 8 -channel chip, named HEROCV1 (HElium-3 ReadOut Circuits), achieves an input dynamic range from 10 fC to 1 pC, with a counting rate of up to 100 kHz. The equivalent noise charge (ENC) measurement is 1297e-@15pF, and the power consumption is less than 9.9 mW per channel. The ASIC has replaced the traditional front-end circuit based on discrete components, fundamentally addressing the issue of front-end power consumption and enabling the entire detector system to operate in a vacuum environment.
更多
查看译文
关键词
Front-end electronics,ASIC,Neutron detector,Position sensitive detector