Capacitive model of enzyme-modified field-effect biosensors: Impact of enzyme coverage

Tobias Karschuck,Arshak Poghossian, Joey Ser, Astghik Tsokolakyan,Stefan Achtsnicht,Patrick Wagner, Michael J. Schöning

Sensors and Actuators B: Chemical(2024)

引用 0|浏览2
暂无评分
摘要
Electrolyte-insulator-semiconductor capacitors (EISCAP) belong to field-effect sensors having an attractive transducer architecture for constructing various biochemical sensors. In this study, a capacitive model of enzyme-modified EISCAPs has been developed and the impact of the surface coverage of immobilized enzymes on its capacitance-voltage and constant-capacitance characteristics was studied theoretically and experimentally. The used multicell arrangement enables a multiplexed electrochemical characterization of up to sixteen EISCAPs. Different enzyme coverages have been achieved by means of parallel electrical connection of bare and enzyme-covered single EISCAPs in diverse combinations. As predicted by the model, with increasing the enzyme coverage, both the shift of capacitance-voltage curves and the amplitude of the constant-capacitance signal increase, resulting in an enhancement of analyte sensitivity of the EISCAP biosensor. In addition, the capability of the multicell arrangement with multi-enzyme covered EISCAPs for sequentially detecting multianalytes (penicillin and urea) utilizing the enzymes penicillinase and urease has been experimentally demonstrated and discussed.
更多
查看译文
关键词
Field-effect biosensor,Capacitive model,Enzyme coverage,Multianalyte detection,Penicillin,Urea
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要