Mirrored port placement for robotic radical prostatectomy with the Hugo RAS? System: initial experience

Alessandro Veccia, Sarah Malandra, Francesca Montanaro, Greta Pettenuzzo,Vincenzo De Marco,Alessandro Antonelli

MINERVA UROLOGY AND NEPHROLOGY(2024)

引用 0|浏览0
暂无评分
摘要
Herein we report our first experience with Hugo RAS (TM) proposing a mirrored approach with different angles. Two experienced surgeons performed 10 prostatectomies (six with the standard approach and four with the mirrored one). The median docking time was 12.5 (IQR 12-15) vs. 13.5 (IQR 12-20) minutes. The median console time was 229 (174-245) vs. 172 (IQR 164-191) minutes. None of the procedures required conversion to open surgery. The study proves the versatility of the Hugo RASTM to perform robot-assisted radical prostatectomy with two different docking angles and might be useful for novel users to adopt the preferred approach.
更多
查看译文
关键词
Robotic surgical procedures,Prostatectomy,Prostatic neoplasms.