Mirrored port placement for robotic radical prostatectomy with the Hugo RAS? System: initial experience
MINERVA UROLOGY AND NEPHROLOGY(2024)
摘要
Herein we report our first experience with Hugo RAS (TM) proposing a mirrored approach with different angles. Two experienced surgeons performed 10 prostatectomies (six with the standard approach and four with the mirrored one). The median docking time was 12.5 (IQR 12-15) vs. 13.5 (IQR 12-20) minutes. The median console time was 229 (174-245) vs. 172 (IQR 164-191) minutes. None of the procedures required conversion to open surgery. The study proves the versatility of the Hugo RASTM to perform robot-assisted radical prostatectomy with two different docking angles and might be useful for novel users to adopt the preferred approach.
更多查看译文
关键词
Robotic surgical procedures,Prostatectomy,Prostatic neoplasms.
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要