Characterization of the IDENTIFY? surface scanning system for radiation therapy setup on a closed-bore linac

Janita Dekker, Sander van het Schip, Marion Essers, Mariska de Smet,Martijn Kusters, Willy de Kruijf

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS(2024)

引用 0|浏览0
暂无评分
摘要
Purpose: In radiation therapy, surface guidance can be used for patient setup and intra-fraction motion monitoring. The surface guided radiation therapy (SGRT) system from Varian Medical systems, IDENTIFYTM, consists of three pods, including cameras and a random pattern projector, mounted on the ceiling. The information captured by the cameras is used to make a reconstruction of the surface. The aim of the study was to assess the technical performance of this SGRT system on a closed-bore linac. Methods: Phantom measurements were performed to assess the accuracy, precision, reproducibility and temporal stability of the system, both in align and in load position. Translations of the phantoms in lateral, longitudinal, and vertical direction, and rotations around three axes (pitch, roll and yaw) were performed with an accurate, in-house built, positioning stage. Different phantom geometries and different surface colors were used, and various ambient light intensities were tested. Results: The accuracy of the IDENTIFYTM system at the closed-bore linac was 0.07 mm and 0.07 degrees at load position, and 0.06 mm and 0.01 degrees at align position for the white head phantom. The precision was 0.02 mm and 0.02 degrees in load position, and 0.01 mm and 0.02 degrees in align position. The accuracy for the Penta-Guide phantom was comparable to the white head phantom, with 0.06 mm and 0.01 degrees in align position. The system was slightly less accurate for translations of the CIRS lung phantom in align position (0.20 mm, 0.05 degrees). Reproducibility measurements showed a variation of 0.02 mm in load position. Regarding the temporal stability, the maximum drift over 30 min was 0.33 mm and 0.02 degrees in load position. No effect of ambient light level on the accuracy of the IDENTIFYTM system was observed. Regarding different surface colors, the accuracy of the system for a black phantom was slightly worse compared to a white surface, but not clinical relevant. Conclusion: The IDENTIFYTM system can adequately be used for motion monitoring on a closed-bore linac with submillimeter accuracy. The results of the performed measurements meet the clinical requirements described in the guidelines of the AAPM and the ESTRO.
更多
查看译文
关键词
accuracy,phantom measurements,surface guided radiation therapy,technical performance
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要