Reference-free X-Ray Fluorescence Analysis with a Micrometer-Sized Incident Beam

Nanotechnology(2024)

Cited 0|Views11
No score
Abstract
Spatially resolved x-ray fluorescence (XRF) based analysis employing incident beam sizes in the low micrometer range (μXRF) is widely used to study lateral composition changes of various types of microstructured samples. However, up to now the quantitative analysis of such experimental datasets could only be realized employing adequate calibration or reference specimen. In this work, we extent the applicability of the so-called reference-free XRF approach to enable reference-freeμXRF analysis. Here, no calibration specimen are needed in order to derive a quantitative and position sensitive composition of the sample of interest. The necessary instrumental steps to realize reference-freeμXRF are explained and a validation of ref.-freeμXRF against ref.-free standard XRF is performed employing laterally homogeneous samples. Finally, an application example from semiconductor research is shown, where the lateral sample features require the usage of ref.-freeμXRF for quantitative analysis.
More
Translated text
Key words
x-ray fluorescence,micro focussed XRF,reference-free analysis,thin-film characterization,nanostructure characterization