Beam Flux Measurement Using Photon Activation Analysis Method at SLEGS

chinaxiv(2024)

引用 0|浏览2
暂无评分
摘要
The Shanghai Laser Electron Gamma Source(SLEGS) has delivered a quasi-monochromatic, continuously energy-tunable gamma-ray beam. Based on the photon activation analysis(PAA) method, SLEGS has built and developed a photon activation analysis platform, including online activation and offline low-background HPGe detector measurement systems, as an alternative to direct measurement methods and a cross-tests at low throughput. Due to the short half-lives spanning from minutes to days and characteristics such as ease of fabrication, cost-effectiveness, and stability, gold (197Au) and zinc (64Zn) emerge as favorable activation targets in the gamma-ray beam flux monitor. Notably, their utility shows a multitude of advantages in monitoring the gamma-ray beam flux typically 105 photons/s with the energy of 13.16 to 19.08 MeV under the condition of the 3 mm coarse collimator. Especially in high-flux $\gamma$-ray beam experiments can be well applied.
更多
查看译文
关键词
SLEGS,Laser Compton Scattering,Beam Flux,Photon Activation Analysis