Patlak Slope versus Standardized Uptake Value Image Quality in an Oncologic PET/CT Population: A Prospective Cross-Sectional Study

Diagnostics(2024)

引用 0|浏览2
暂无评分
摘要
Patlak slope (PS) images have the potential to improve lesion conspicuity compared with standardized uptake value (SUV) images but may be more artifact-prone. This study compared PS versus SUV image quality and hepatic tumor-to-background ratios (TBRs) at matched time points. Early and late SUV and PS images were reconstructed from dynamic positron emission tomography (PET) data. Two independent, blinded readers scored image quality metrics (a four-point Likert scale) and counted tracer-avid lesions. Hepatic lesions and parenchyma were segmented and quantitatively analyzed. Differences were assessed via the Wilcoxon signed-rank test (alpha, 0.05). Forty-three subjects were included. For overall quality and lesion detection, early PS images were significantly inferior to other reconstructions. For overall quality, late PS images (reader 1 [R1]: 3.95, reader 2 [R2]: 3.95) were similar (p > 0.05) to early SUV images (R1: 3.88, R2: 3.84) but slightly superior (p ≤ 0.002) to late SUV images (R1: 2.97, R2: 3.44). For lesion detection, late PS images were slightly inferior to late SUV images (R1 only) but slightly superior to early SUV images (both readers). PS-based TBRs were significantly higher than SUV-based TBRs at the early time point, with opposite findings at the late time point. In conclusion, late PS images are similar to early/late SUV images in image quality and lesion detection; the superiority of SUV versus PS hepatic TBRs is time-dependent.