A Compact Start Time Counter Using Plastic Scintillators Readout with MPPC Arrays for the WASA-FRS HypHI Experiment

Nuclear Instruments and Methods in Physics Research Section A Accelerators, Spectrometers, Detectors and Associated Equipment(2024)

引用 0|浏览19
暂无评分
摘要
We have developed a compact detector for measuring beam particles using plastic scintillators readout through Multi-Pixel Photon Counters, which is employed for hypernuclear measurements in the WASA-FRS experiment at GSI. The Time-of-Flight resolution of the newly-developed detector has been investigated in relation to the overvoltage with respect to the breakdown voltage, a maximum counting rate of approximately 3×106/s per segment, and a maximum beam charge of Z = 6. The evaluated Time-of-Flight resolutions between the neighboring segments of the detector range from 44.6±1.3 ps to 100.3±3.6 ps (σ) depending on the segment, overvoltage values, and beam intensity. It is also observed that the Time-of-Flight resolution is inversely correlated to the beam atomic charge (Z).