Near-infrared Imaging in Orthodontic Intraoral Scanners for Early Interproximal Caries Detection

Adam Hoxie, Abhinaya Perumbedu, Parth Patel,Jialiu Xie,Kelly Mitchell,Angela Broome, Adalberto Bastos Vasconcellos,Apoena Aguiar Ribeiro

American journal of orthodontics and dentofacial orthopedics(2024)

引用 0|浏览2
暂无评分
摘要
IntroductionIntraoral scanners commonly used in orthodontic offices now offer near-infrared imaging (NIRI) technology, advertised as a screening tool to identify interproximal caries. This study aimed to evaluate the reliability and validity of NIRI detection of interproximal carious lesions in a common intraoral scanner (iTero Element 5D; Align Technology, San Jose, Calif) with and without bitewing radiograph complement, compared with a microcomputed tomography (micro-CT) reference standard.MethodsExtracted human posterior teeth (premolars and molars) were selected for early (noncavitated) interproximal carious lesions (n = 39) and sound control surfaces (n = 47). The teeth were scanned via micro-CT for evaluation by 2 blinded evaluators using consensus scoring. The teeth were mounted to simulate anatomic interproximal contacts and underwent a NIRI scan using iTero Element 5D and bitewing radiographs. Two trained, calibrated examiners independently evaluated (1) near-infrared images alone with clinical photograph, (2) bitewing radiograph alone with clinical photograph, and (3) near-infrared images with bitewing radiograph and clinical photograph in combination, after at least a 10-day washout period between each evaluation.ResultsInterrater reliability was highest for NIRI alone (k = 0.533) compared with bitewing radiograph alone (k = 0.176) or in combination (k = 0.256). NIRI alone showed high specificity (0.83-0.96) and moderate sensitivity (0.42-0.63) compared with a micro-CT reference standard. Dentin lesions were significantly more reliably detected than enamel lesions.ConclusionsAfter rigorous training and calibration, NIRI can be used with moderate reliability, high specificity, and moderate sensitivity to detect noncavitated interproximal carious lesions.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要