Investigation of triaxial cables and microdetectors in small field dosimetry

BIOMEDICAL PHYSICS & ENGINEERING EXPRESS(2024)

引用 0|浏览4
暂无评分
摘要
Background. Small field dosimetry presents unique challenges with source occlusion, lateral charged particle equilibrium and detector size. As detector volume decreases, signal strength declines while noise increases, deteriorating the signal-to-noise ratio (SNR). This issue may be compounded by triaxial cables connecting detectors to electrometers. However, effects of cables, critical for precision dosimetry, are often overlooked. There is a need to evaluate triaxial cable and detector impacts on SNR in small fields. The purpose of this study is to evaluate the influence of triaxial cables and microdetectors on signal-to-noise ratios in small-field dosimetry. This study also aims to establish the importance of cable quality assurance for measurement accuracy. Methods. Six 9.1 m length triaxial cables from different manufacturers were tested with six microdetectors (microDiamond, PinPoint, EDGE, Plastic scintillator, microSilicon, SRS-Diode). A 6 MV photon beam (TrueBeam) was used, with a water phantom at 5 cm depth with 0.5 x 0.5 cm2 to 10 x 10 cm2 fields at 600 MU min-1. Readings were acquired using cable-detector permutations with a dedicated electrometer (except the scintillator which has its own). Cables had differing connector types, conductor materials, insulation, and diameters. Detectors had various sensitive volumes, materials, typical signals, and bias voltages. Results. Normalized field output correction factors (FOFs) relative differences of 13.4% and 4.6% between the highest and lowest values across triaxial cables for 0.5 x 0.5 cm2 and 1 x 1 cm2 fields, respectively. The maximum difference in FOF between any cable-detector combinations was 0.2% for the smallest field size. No consistent FOF trend was observed across all detectors when increasing cable diameter. Additionally, the non-normalized FOF differences of 0.9% and 0.3% were observed between cables for 0.5 x 0.5 cm2 and 1 x 1 cm2 fields, respectively. Conclusions. Regular triaxial cable quality assurance is critical for precision small field dosimetry. A national protocol is needed to standardize cable evaluations/calibrations, particularly for small signals (更多
查看译文
关键词
small field dosimetry,triaxial cable,microdetectors,SNR,QA
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要