BACKGROUND Right atrial (RA) remodeling is closely related to disease severity and outcomes in pulmonary arterial hypertension (PAH), but the prognostic value of coupling RA structure with mechanical performance remains uncertain. OBJECTIVES To evaluate the clinical correlates, reproducibility, prognostic value, and incremental prognostic information of a three-dimensional right atrial coupling index (RACI) in PAH. METHODS In this single-center bidirectional cohort, 158 treatment-naïve patients with PAH underwent right heart catheterization and echocardiography. RACI was calculated as the three-dimensional right atrial volume index divided by tissue Doppler a' velocity. The primary outcome was all-cause death or hospitalization for worsening right heart failure. Prognostic associations were assessed using multivariable Cox regression, Kaplan-Meier analysis, time-dependent ROC analysis, nested models, and bootstrap internal validation. RESULTS During follow-up, 27 patients died and 36 were hospitalized for right heart failure. Each 2-unit increase in RACI was independently associated with the composite outcome (HR 1.21, 95% CI 1.04-1.41; P = 0.016), while the cohort-derived cutoff of 3.84 was associated with higher risk (HR 4.09, 95% CI 1.95-8.58; P < 0.001). Higher RACI correlated with larger RA size, higher mRAP/NT-proBNP, and worse RV function. Adding RACI to COMPERA 2.0 and mRAP yielded only a modest increase in C-index (ΔC=0.011, 95% CI -0.008-0.052; P = 0.456), although model fit improved. Intra- and interobserver ICCs were 0.98 and 0.96, respectively. CONCLUSION RACI was reproducible and independently associated with adverse outcomes in PAH, providing complementary prognostic information that requires prospective multicenter validation.
Background Body weight guides decongestion in acute decompensated heart failure and is frequently used as a trial outcome, but scale performance in routine care is insufficiently characterized. Objectives To assess how closely household and medical scales measured known reference loads, how much measurements varied between scales, and how consistently scales performed during repeated measurements and across measurement days. Methods We tested 37 digital scales (29 household and 8 medical) at two Danish hospitals using reference loads from 25 to 125 kg. Each scale underwent three measurements per load on two days. Results Household scales showed modest mean measurement error at most reference loads, but variability between devices increased and repeated and day-to-day measurements became less consistent as load increased. At 125 kg, median absolute measurement error was 1.13 kg on Day 1 and 1.43 kg on Day 2, while between-scale standard deviations (SDs) were 2.30 kg and 6.27 kg, respectively. Medical scales showed mean errors of 0.00–0.25 kg, between-scale SDs of 0.00–0.12 kg, consistently high repeatability, and small day-to-day differences. Conclusion Among the scales tested, household scales showed modest mean measurement error but substantial between-scale variability and poorer repeatability at higher loads, whereas medical scales performed consistently. These findings support preferential use of medical scales in hospitals and use of the same device for serial measurements when household scales are used.