Imaging phantoms are used to calibrate and validate the performance of medical computed tomography (CT) systems. Many new materials developed for three-dimensional (3D) printing processes may be useful in the direct printing or casting of biomimetic and geometrically accurate CT and X-ray phantoms. The X-ray linear attenuation coefficients of polymer samples were measured to discover materials for use as tissue mimics in phantoms. This study included a cohort of polymer compounds that were tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed with a commercial micro-CT imaging system from 40 kVp to 140 kVp. The X-ray linear coefficients of the samples and human tissues were plotted with error bars to allow the reader to identify suitable mimics. The X-ray linear attenuation coefficients of the tested material samples spanned a wide range of values, with a small number of them overlapping established human tissue mimic values. Twenty 3D printer materials and one castable polyurethane tracked nylon and polymethyl methacrylate (PMMA) as established X-ray mimics for fat. Five 3D printer materials tracked water as an established X-ray mimic for muscle.
Imaging phantoms are used to calibrate and validate the performance of magnetic resonance imaging (MRI) systems. Many new materials have been developed for additive manufacturing (three-dimensional [3D] printing) processes that may be useful in the direct printing or casting of dimensionally accurate, anatomically accurate, patient-specific, and/or biomimetic MRI phantoms. The T1, T2, and T2* spin relaxation times of polymer samples were tested to discover materials for use as tissue mimics and structures in MRI phantoms. This study included a cohort of polymer compounds that was tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed at 3 T using inversion recovery, spin echo, and gradient echo sequences for T1, T2, and T2*, respectively. T1, T2, and T2* values were plotted with error bars to allow the reader to assess how well a polymer matches a tissue for a specific application. A correlation was performed between T1, T2, T2* values and material density, elongation, tensile strength, and hardness. Two silicones, SI_XP-643 and SI_P-45, may be usable mimics for reported liver values; one silicone, SI_XP-643, may be a useful mimic for muscle; one silicone, SI_XP-738, may be a useful mimic for white matter; and four silicones, SI_P-15, SI_GI-1000, SI_GI-1040, and SI_GI-1110, may be usable mimics for spinal cord. Elongation correlated to T2 (p = 0.0007), tensile strength correlated to T1 (p = 0.002), T2 (p = 0.0003), and T2* (p = 0.003). The 80 samples not providing measurable signal with T1, T2, T2* relaxation values too short to measure with the standard sequences, may be useful for MRI-invisible fixturing and medical devices at 3 T.
PURPOSE:This study outlines the design and fabrication techniques for two portal vein flow phantoms.METHODS:A materials study was performed as a precursor to this phantom fabrication effort and the desired material properties are restated for continuity. A three-dimensional portal vein pattern was created from the Visual Human database. The portal vein pattern was used to fabricate two flow phantoms by different methods with identical interior surface geometry using computer aided design software tools and rapid prototyping techniques. One portal flow phantom was fabricated within a solid block of clear silicone for use on a table with Ultrasound or within medical imaging systems such as MRI, CT, PET, or SPECT. The other portal flow phantom was fabricated as a thin walled tubular latex structure for use in water tanks with Ultrasound imaging. Both phantoms were evaluated for usability and durability.RESULTS:Both phantoms were fabricated successfully and passed durability criteria for flow testing in the next project phase.CONCLUSIONS:The fabrication methods and materials employed for the study yielded durable portal vein phantoms.
PURPOSE:This study investigated the ultrasound, MRI, and CT imaging characteristics of several industrial casting and molding compounds as a precursor to the future development of durable and anatomically correct flow phantoms.METHODS:A set of usability and performance criteria was established for a proposed phantom design capable of supporting liquid flow during imaging. A literature search was conducted to identify the materials and methods previously used in phantom fabrication. A database of human tissue and casting material properties was compiled to facilitate the selection of appropriate materials for testing. Several industrial casting materials were selected, procured, and used to fabricate test samples that were imaged with ultrasound, MRI, and CT.RESULTS:Five silicones and one polyurethane were selected for testing. Samples of all materials were successfully fabricated. All imaging modalities were able to discriminate between the materials tested. Ultrasound testing showed that three of the silicones could be imaged to a depth of at least 2.5 cm (1 in.). The RP-6400 polyurethane exhibited excellent contrast and edge detail for MRI phantoms and appears to be an excellent water reference for CT applications. The 10T and 27T silicones appear to be usable water references for MRI imaging.CONCLUSIONS:Based on study data and the stated selection criteria, the P-4 silicone provided sufficient material contrast to water and edge detail for use across all imaging modalities with the benefits of availability, low cost, dimensional stability, nontoxic, nonflammable, durable, cleanable, and optical clarity. The physical and imaging differences of the materials documented in this study may be useful for other applications.