Organic and elemental carbon (OC and EC) are operationally-defined by the measurement process, so long-term trends may be interrupted with instrumentation changes. A modification to the U.S. IMPROVE carbon analysis protocol and hardware is examined that replaces the 633 nm laser light used for OC charring adjustments with seven wavelengths ranging from 405 to 980 nm, including one at 635 nm. Reflectance (R) and Transmittance (T) values for each wavelength are made traceable to primary standards through transfer standards consisting of a range of aerosol deposits on filter media similar to that of the analyzed samples. R and T values are assigned to these filters using a UV/VIS spectrometer calibrated with these standards. Using ambient and source (e.g., diesel exhaust, flaming biomass, and smoldering biomass) samples, it is demonstrated that R and T calibration is independent of the sample type. Total carbon (TC), OC, and EC comparisons with the earlier hardware design for urban- and non-urban samples demonstrate equivalence, within precisions derived from replicate analyses, for the 633 nm and 635 nm wavelengths. Several uses of the additional multiwavelength information are identified, including: 1) ground-truthing of multi-spectral remote sensors; 2) improving estimates of the Earth's radiation balance; 3) associating specific organic compounds with their light absorption properties; and 4) appropriating sources of black and brown carbon.
The replacement of the Desert Research Institute (DRI) model 2001 with model 2015 thermal/optical analyzers (TOAs) results in continuity of the long-term organic carbon (OC) and elemental carbon (EC) database, and it adds optical information with no additional carbon analysis effort. The value of multiwavelength light attenuation is that light absorption due to black carbon (BC) can be separated from that of brown carbon (BrC), with subsequent attribution to known sources such as biomass burning and secondary organic aerosols. There is evidence of filter loading effects for the 25% of all samples with the highest EC concentrations based on the ratio of light attenuation to EC. Loading corrections similar to those used for the seven-wavelength aethalometer need to be investigated. On average, nonurban Interagency Monitoring of PROtected Visual Environments (IMPROVE) samples show higher BrC fractions of short-wavelength absorption than urban Chemical Speciation Network (CSN) samples, owing to greater influence from biomass burning and aged aerosols, as well as to higher primary BC contributions from engine exhaust at urban sites. Sequential samples taken during an Everglades National Park wildfire demonstrate the evolution from flaming to smoldering combustion, with the BrC fraction increasing as smoldering begins to dominate the fire event. Implications: The inclusion of seven wavelengths in thermal/optical carbon analysis of speciated PM2.5 (particulate matter with an aerodynamic diameter ≤2.5 μm) samples allows contributions from biomass burning and secondary organic aerosols to be estimated. This separation is useful for evaluating control strategy effectiveness, identifying exceptional events, and determining natural visibility conditions.
Accurate, precise, and valid organic and elemental carbon (OC and EC, respectively) measurements require more effort than the routine analysis of ambient aerosol and source samples. This paper documents the quality assurance (QA) and quality control (QC) procedures that should be implemented to ensure consistency of OC and EC measurements. Prior to field sampling, the appropriate filter substrate must be selected and tested for sampling effectiveness. Unexposed filters are pre-fired to remove contaminants and acceptance tested. After sampling, filters must be stored in the laboratory in clean, labeled containers under refrigeration (<4 °C) to minimize loss of semi-volatile OC. QA activities include participation in laboratory accreditation programs, external system audits, and interlaboratory comparisons. For thermal/optical carbon analyses, periodic QC tests include calibration of the flame ionization detector with different types of carbon standards, thermogram inspection, replicate analyses, quantification of trace oxygen concentrations (<100 ppmv) in the helium atmosphere, and calibration of the sample temperature sensor. These established QA/QC procedures are applicable to aerosol sampling and analysis for carbon and other chemical components.
Summary of filter handling procedures for different long-term networks. 2-14 Table 2-3. Collocated PM 2.5 speciation data for the IMPROVE network and STN/CSN.........2-15 Table 2-4. Examples of organic carbon (OC) database differences among air quality networks 2-16 Table 2-5. Tests for diesel and levoglucosan-doping experiments. 2-17 Table 2-6. Comparison of the IMPROVE, IMPROVE_A, and STN/CSN thermal/optical carbon analysis protocols 2-18 Table 3-1. Comparison of average field blank levels (± standard deviation) among the IMPROVE, STN/CSN, and SEARCH networks for the period from 1/1/2005 through 12/31/2006 3-28 Table 3-2. Average field blank concentration by carbon fraction at the 181 sites in the IMPROVE network from 1/1/2005 to 12/31/2006 3-29 Table 3-3. Average field blank concentrations by thermal carbon fractions at the 239 sites in STN/CSN for the period from 1/1/2005 through 12/31/2006. 3-35 Table 3-4. Average field blank concentrations by carbon fractions at the eight sites in the SEARCH network for the period from 1/1/2005 through 12/31/2006 3-45 Table 3-5. Average blank concentration by PM
Thermally derived carbon fractions including organic carbon (OC) and elemental carbon (EC) have been reported for the U.S. Interagency Monitoring of PROtected Visual Environments (IMPROVE) network since 1987 and have been found useful in source apportionment studies and to evaluate quartz-fiber filter adsorption of organic vapors. The IMPROVE_A temperature protocol defines temperature plateaus for thermally derived carbon fractions of 140 degrees C for OC1, 280 degrees C for OC2, 480 degrees C for OC3, and 580 degrees C for OC4 in a helium (He) carrier gas and 580 degrees C for EC1, 740 degrees C for EC2, and 840 degrees C for EC3 in a 98% He/2% oxygen (O2) carrier gas. These temperatures differ from those used previously because new hardware used for the IMPROVE thermal/optical reflectance (IMPROVE_TOR) protocol better represents the sample temperature than did the old hardware. A newly developed temperature calibration method demonstrates that these temperatures better represent sample temperatures in the older units used to quantify IMPROVE carbon fractions from 1987 through 2004. Only the thermal fractions are affected by changes in temperature. The OC and EC by TOR are insensitive to the change in temperature protocol, and therefore the long-term consistency of the IMPROVE database is conserved. A method to detect small quantities of O2 in the pure He carrier gas shows that O2 levels above 100 ppmv also affect the comparability of thermal carbon fractions but have little effect on the IMPROVE_TOR split between OC and EC.
Thermal/optical methods have been widely used for quantifying total carbon (TC), organic carbon (OC), and elemental carbon (EC) in ambient and source particulate samples. Thermally defined carbon fractions have been used for source identification. Temperature precision in thermal carbon analysis is critical to the allocation of carbon fractions. The sample temperature is determined by a thermocouple, which is usually located in the oven near the sample. Sample and thermocouple temperature may differ owing to different thermal properties between the sample filter punch and the thermocouple, or inhomogeneities in the heating zone. Quick-drying temperature-indicating liquids (Tempil Inc., South Plainfield, NJ) of different liquefying points are used as temperature calibration standards. These consist of chemicals that change their appearance at specific temperatures and can be optically monitored to determine the sample temperature. Temperature measures were evaluated for three different models of carbon analyzers. Sample temperatures were found to differ from sensor temperatures by 10 to 50°C. Temperature biases of 14 to 22°C during thermal analysis were found to change carbon fraction measurements. The temperature indicators allow calibration curves to be constructed that relate the sample temperature to the temperature measured by a thermocouple.