Glycols, specifically ethylene glycol and 1,2-propanediol, are some of the major organic compounds found in the humidity condensate samples collected on the International Space Station. The current analytical method for glycols is a GC/MS method with direct sample injection. This method is simple and fast, but it is not very sensitive. Reporting limits for ethylene glycol and 1,2-propanediol are only 1 ppm. A much more sensitive GC/FID method was developed, in which glycols were derivatized with benzoyl chloride for 10 minutes before being extracted with hexane. Using 1,3-propanediol as an internal standard, the detection limits for the GC/FID method was determined to be 50 ppb and the analysis only takes 7 minutes. Data from the GC/MS and the new GC/FID methods shows excellent agreement with each other. Factors affecting the sensitivity, including sample volume, NaOH concentration and volume, volume of benzoyl chloride, reaction time and temperature, were investigated. Interferences during derivatization and possible method to reduce interferences were also investigated.
Maintaining a safe supply of potable water is of utmost importance when preparing for long-duration spaceflight missions, with the minimization of microbial growth being one major aspect. While biocides, such as ionic silver, historically have been used for microbial control in spaceflight, their effectiveness is sometimes limited due to surface reactions with the materials of the storage containers that reduce their concentrations below the effective range. For the Multi-Purpose Crew Vehicle, the primary wetted materials of the water storage system are stainless steel and a titanium alloy, and ionic silver has been chosen to serve as the biocide. As an attempt to understand what processes might reduce the known losses of silver, different treatment processes were attempted and samples of the wetted materials were tested, individually and together, to determine the relative loss of biocide under representative surface area-to-volume ratios. The results of testing presented here showed that the materials could be treated by a nitric acid rinse or a high-concentration silver spike to reduce the loss of silver and bacterial growth. It was also found that the minimum biocidal concentration could be maintained for over 28 days. These results have pointed to approaches that could be used to successfully maintain silver in spacecraft water systems for long-duration missions.
Dimethylsilanediol (DMSD) has been identified as a problematic organic on ISS. This contaminant was initially identified in the Water Processor Assembly (WPA) product water in 2010 by the Total Organic Carbon Analyzer (TOCA). DMSD is not a crew health hazard at the levels observed in the product water, but it may degrade the performance of the Oxygen Generation System (OGS) which uses the WPA product water for electrolysis and does impact the effective operation of the WPA catalytic reactor. To mitigate these impacts, early replacement of the Multifiltration Beds in the WPA is required. An investigation has determined that the decomposition of atmospheric polydimethylsiloxanes (PDMSs) is the primary source of DMSD in the condensate. PDMSs are prevalent on ISS from a variety of sources, including crew hygiene products, adhesives, caulks, lubricants, and various nonmetallics. These PDMSs also contribute to degradation of the CHX hydrophilic coating, rendering it hydrophobic and therefore affecting its ability to transmit water to the condensate bus. In addition, literature research has determined that PDMSs are likely oxidized to DMSD in the atmosphere when exposed to hydroxyl radicals in the ISS atmosphere. To address these mechanisms, filters have been developed for removal of PDMSs from the ISS atmosphere. However, ongoing analysis indicates a significant reduction in atmospheric PDMSs is required to achieve a measurable reduction of DMSD in the condensate. As a result, additional measures are being pursued to mitigate this issue. First, credible sources are being investigated to quantity to the extent possible the significant sources of PDMSs and identify sources that can be reasonably removed from ISS. Second, a Reverse Osmosis technology is being investigated as an alternate means for removing DMSD from the condensate. This paper summarizes the current status of the overall effort to mitigate DMSD in the US condensate.
The International Space Station (ISS) Environmental Control and Life Support System (ECLSS) provides a working environment for six crewmembers through atmosphere revitalization and water recovery systems. In the last year, elevated ethanol levels have presented a unique challenge for the ISS ECLSS. Ethanol is monitored on the ISS by the Air Quality Monitor (AQM). The source of this increase is currently unknown. This paper documents the credible sources for the increased ethanol concentration, the monitoring provided by the AQM, and the impact on the atmosphere revitalization and water recovery systems.
When preparing for long-duration spaceflight missions, maintaining a safe supply of potable water is of the utmost importance. One major aspect of that is ensuring that microbial growth is minimized. Historically, this challenge has been addressed through the use of biocides. When using biocides, the choice of materials for the storage containers is important, because surface reactions can reduce biocide concentrations below their effective range. In the water storage system baselined for the Orion vehicle, the primary wetted materials are stainless steel (316 L) and a titanium alloy (Ti6Al4V). Previous testing with these materials has shown that the biocide selected for use in the system (ionic silver) will plate out rapidly upon initial wetting of the system. One potential approach for maintaining an adequate biocide concentration is to spike the water supply with high levels of biocide in an attempt to passivate the surface. To evaluate this hypothesis, samples of the wetted materials were tested individually and together to determine the relative loss of biocide under representative surface area-to-volume ratios after 24 hours. Additionally, we have analyzed the efficacy of disinfecting a system containing these materials by measuring reductions in bacterial counts in the same test conditions. Preliminary results indicate that the use of titanium, either individually or in combination with stainless steel, can result in over 95% loss of biocide, while less than 5% is lost when using stainless steel. In bacterial testing, viable organisms were recovered from samples exposed to the titanium coupons after 24 hours. By comparison, no organisms were recovered from the test vessels containing only stainless steel. These results indicate that titanium, while possessing some favorable attributes, may pose additional challenges when used in water storage tanks with ionic silver biocide.
Low molecular weight polar organics are commonly observed in spacecraft environments. Increasing concentrations of one or more of these contaminants can negatively impact Environmental Control and Life Support (ECLS) systems and/or the health of crew members, posing potential risks to the success of manned space missions. Ambient plasma ionization mass spectrometry (MS) is finding effective use as part of the analytical methodologies being tested for next-generation space module environmental analysis. However, ambient ionization methods employing atmospheric plasmas typically require relatively high operation voltages and power, thus limiting their applicability in combination with fieldable mass spectrometers. In this work, we investigate the use of a low power microplasma device in the microhollow cathode discharge (MHCD) configuration for the analysis of polar organics encountered in space missions. A metal-insulator-metal (MIM) structure with molybdenum foil disc electrodes and a mica insulator was used to form a 300 μm diameter plasma discharge cavity. We demonstrate the application of these MIM microplasmas as part of a versatile miniature ion source for the analysis of typical volatile contaminants found in the International Space Station (ISS) environment, highlighting their advantages as low cost and simple analytical devices.
2015 marked the 15th anniversary of continuous human presence on board the International Space Station. During the past year crew members from Expeditions 42-46, including two participating in a one-year mission, continued to rely on reclaimed water as their primary source of potable water. This paper presents and discusses results from chemical analyses performed on ISS water samples returned in 2015. Since the U.S. water processor assembly (WPA) became operational in 2008, there have been 5 instances of organic contaminants breaking through the treatment process. On each occasion, the breakthrough was signaled by an increase in the total organic carbon (TOC) concentration in the product water measured by the onboard TOC analyzer (TOCA). Although the most recent TOC rise in 2015 was not unexpected, it was the first time where dimethylsilanediol (DMSD) was not the primary compound responsible for the increase. Results from ground analysis of a product water sample collected in June of 2015 and returned on Soyuz 41 showed that DMSD only accounted for 10% of the measured TOC. After considerable laboratory investigation, the compound responsible for the majority of the TOC was identified as monomethysilanetriol (MMST). MMST is a low-toxicity compound that is structurally similar to DMSD.
The water produced by the U.S. Water Processor Assembly (WPA) on the International Space Station had virtually undetectable levels of total organic carbon (TOC) during the first 18 months of operation. In 2010, the TOC in the WPA product water rose to approx. 2 mg/L. The maximum allowable TOC concentration in the product water is 3.0 mg/L. Ground analysis indicated that > 90% of this increase was due to dimethylsilanediol (DMSD). Since then, recurring increases in TOC attributed to DMSD have occurred every 12-18 months. In 2015, a new compound was determined to be responsible for the TOC increase. This new contaminant was identified as monomethylsilanetriol (MMST). This poster details the efforts to identify and quantify DMSD and MMST. DMSD was initially discovered as a stray peak in a routine GC/MS analysis used to measure glycols. This method was not suitable for routine analysis of DMSD because of extreme carryover problems. MMST was discovered and identified using direct infusion electrospray ionization (ESI) mass spectrometry. The identification was confirmed by purchasing an MMST standard and analyzing it by HPLC with refractive index (RI) detection. An LC/MS method has been developed, based on the LC/RI method that quantifies both DMSD and MMST in a 14-minute isocratic run. This method employs atmospheric pressure chemical ionization (APCI) and a single quadrupole mass spectrometer operated in negative mode with selected ion monitoring.
In the history of manned spaceflight, environmental monitoring has relied heavily on archival sampling. However, with the construction of the International Space Station (ISS) and the subsequent extension in mission duration up to one year, an enhanced, real-time method for environmental monitoring is necessary. The station air is currently monitored for trace volatile organic compounds (VOCs) using gas chromatography-differential mobility spectrometry (GC-DMS) via the Air Quality Monitor (AQM), while water is analyzed to measure total organic carbon and biocide concentrations using the Total Organic Carbon Analyzer (TOCA) and the Colorimetric Water Quality Monitoring Kit (CWQMK), respectively. As mission scenarios extend beyond low Earth orbit, a convergence in analytical instrumentation to analyze both air and water samples is highly desirable. Since the AQM currently provides quantitative, compound-specific information for air samples and many of the targets in air are also common to water, this platform is a logical starting point for developing a multimatrix monitor. Here, we report on the interfacing of an electrothermal vaporization (ETV) sample introduction unit with a ground-based AQM for monitoring target analytes in water. The results show that each of the compounds tested from water have similar GC-DMS parameters as the compounds tested in air. Moreover, the ETV enabled AQM detection of dimethlsilanediol (DMSD), a compound whose analysis had proven challenging using other sample introduction methods. Analysis of authentic ISS water samples using the ETV-AQM showed that DMSD could be successfully quantified, while the concentrations obtained for the other compounds also agreed well with laboratory results.
Environmental monitoring for manned spaceflight has long depended on archival sampling, which was sufficient for short missions. However, the longer mission durations aboard the International Space Station (ISS) have shown that enhanced, real-time monitoring capabilities are necessary in order to protect both the crewmembers and the spacecraft systems. Over the past several years, a number of real-time environmental monitors have been deployed on the ISS. Currently, volatile organic compounds (VOCs) in the station air are monitored by the Air Quality Monitor (AQM), a small, lightweight gas chromatograph-differential mobility spectrometer. For water monitoring, real-time monitors are used for total organic carbon (TOC) and biocide analysis. No information on the actual makeup of the TOC is provided presently, however. An improvement to the current state of environmental monitoring could be realized by modifying a single instrument to analyze both air and water. As the AQM currently provides quantitative, compound-specific information for VOCs in air samples, this instrument provides a logical starting point to evaluate the feasibility of this approach. The major hurdle for this effort lies in the liberation of the target analytes from the water matrix. In this presentation, we will discuss our recent studies, in which an electro-thermal vaporization unit has been interfaced with the AQM to analyze target VOCs at the concentrations at which they are routinely detected in archival water samples from the ISS. We will compare the results of these studies with those obtained from the instrumentation routinely used to analyze archival water samples.
Microbial control in the spacecraft environment is a daunting task, especially in the presence of human crew members. Currently, assessing the potential crew health risk associated with a microbial contamination event requires return of representative environmental samples that are analyzed in a ground-based laboratory. It is therefore not currently possible to quickly identify microbes during spaceflight. This project addresses the unmet need for spaceflight-compatible microbial identification technology. The electrochemical detection and identification platform is expected to provide a sensitive, specific, and rapid sample-to-answer capability for in-flight microbial monitoring that can distinguish between related microorganisms (pathogens and non-pathogens) as well as chemical contaminants. This will dramatically enhance our ability to monitor the spacecraft environment and the health risk to the crew. Further, the project is expected to eliminate the need for sample return while significantly reducing crew time required for detection of multiple targets. Initial work will focus on the optimization of bacterial detection and identification. The platform is designed to release nucleic acids (DNA and RNA) from microorganisms without the use of harmful chemicals. Bacterial DNA or RNA is captured by bacteria-specific probe molecules that are bound to a microelectrode, and that capture event can generate a small change in the electrical current (Lam, et al. 2012. Anal. Chem. 84(1): 21-5.). This current is measured, and a determination is made whether a given microbe is present in the sample analyzed. Chemical detection can be accomplished by directly applying a sample to the microelectrode and measuring the resulting current change. This rapid microbial and chemical detection device is designed to be a low-cost, low-power platform anticipated to be operated independently of an external power source, characteristics optimal for manned spaceflight and areas where power and computing resources are scarce.
Since activation of the Water Processor Assembly (WPA) on the International Space Station (ISS) in November of 2008, there have been three events in which the TOC (Total Organic Carbon) in the product water has increased to approximately 3 mg/L and has subsequently recovered. Analysis of the product water in 2010 identified the primary component of the TOC as dimethylsilanediol (DMSD). An investigation into the fate of DMSD in the WPA ultimately determined that replacement of both Multifiltration (MF) Beds is the solution to recovering product water quality. The MF Beds were designed to ensure that ionic breakthrough occurs before organic breakthrough. However, DMSD saturated both MF Beds in the series, requiring removal and replacement of both MF Beds with significant life remaining. Analysis of the MF Beds determined that the adsorbent was not effectively removing DMSD, trimethylsilanol, various polydimethylsiloxanes, or dimethylsulfone. Coupled with the fact that the current adsorbent is now obsolete, the authors evaluated various media to identify a replacement adsorbent as well as media with greater capacity for these problematic organic contaminants. This paper provides the results and recommendations of this collaborative study.
The development of a direct analysis in real time-mass spectrometry (DART-MS) method and first prototype vaporizer for the detection of low molecular weight (∼30-100 Da) contaminants representative of those detected in water samples from the International Space Station is reported. A temperature-programmable, electro-thermal vaporizer (ETV) was designed, constructed, and evaluated as a sampling interface for DART-MS. The ETV facilitates analysis of water samples with minimum user intervention while maximizing analytical sensitivity and sample throughput. The integrated DART-ETV-MS methodology was evaluated in both positive and negative ion modes to (1) determine experimental conditions suitable for coupling DART with ETV as a sample inlet and ionization platform for time-of-flight MS, (2) to identify analyte response ions, (3) to determine the detection limit and dynamic range for target analyte measurement, and (4) to determine the reproducibility of measurements made with the method when using manual sample introduction into the vaporizer. Nitrogen was used as the DART working gas, and the target analytes chosen for the study were ethyl acetate, acetone, acetaldehyde, ethanol, ethylene glycol, dimethylsilanediol, formaldehyde, isopropanol, methanol, methylethyl ketone, methylsulfone, propylene glycol, and trimethylsilanol.