We present a concept for a combined selection and focus field generator for whole-body magnetic particle imaging (MPI). Predictions for field of view, gradient field strength, slew rate, and power consumption are calculated and discussed with the help of finite-element simulations.
After realizing the worlds' first preclinical magnetic particle imaging (MPI) demonstrator, Philips is now realizing the worlds' first whole-body clinical prototype to prove the feasibility of MPI for clinical imaging. After a brief introduction of the basic MPI imaging process, this contribution presents an overview on the determining factors for key properties, i.e., spatial resolution, acquisition speed, sensitivity, and quantitativeness, and how these properties are influenced by scaling up from preclinical to clinical instrumentation. Furthermore, it is discussed how this scale up affects the physiological compatibility of the method as well as hardware parameters such as power requirements for drive field generation, selection and focus field generation, and the design of the receive chain of the MPI device.
In order to identify suitable operating conditions for future clinical Magnetic Particle Imaging, peripheral nerve stimulation (PNS) and specific absorption rate (SAR) experiments have been performed by exposing volunteers to sinusoidally time-varying magnetic fields along and transverse to the body axis at frequencies from 24 kHz to 162 kHz. The findings show that future clinical MPI can advantageously be performed at elevated drive-field frequencies, with PNS restriction actually relaxed at higher frequencies, and with still acceptable SAR exposure.
Magnetic particle imaging (MPI) is a new medical imaging technique which performs a direct measurement of magnetic nanoparticles, also known as superparamagnetic iron oxide. MPI can acquire quantitative images of the local distribution of the magnetic material with high spatial and temporal resolution. Its sensitivity is well above that of other methods used for the detection and quantification of magnetic materials, for example, magnetic resonance imaging. On the basis of an intravenous injection of magnetic particles, MPI has the potential to play an important role in medical application areas such as cardiovascular, oncology, and also in exploratory fields such as cell labeling and tracking. Here, we present an introduction to the basic function principle of MPI, together with an estimation of the spatial resolution and the detection limit. Furthermore, the above-mentioned medical applications are discussed with respect to an applicability of MPI.
An MR‐electrophysiology (EP) catheter is presented that provides full diagnostic EP functionality and a high level of radiofrequency safety achieved by custom‐designed transmission lines. Highly resistive wires transmit intracardiac electrograms and currents for intracardiac pacing. A transformer cable transmits the localization signal of a tip coil. Specific absorption rate simulations and temperature measurements at 1.5 T demonstrate that a wire resistance > 3 kΩ/m limits dielectric heating to a physiologically irrelevant level. Additional wires do not increase tip specific absorption rate significantly, which is important because some clinical catheters require up to 20 electrodes. It is further demonstrated that radiofrequency‐induced and pacing‐induced resistive heating of the wires is negligible under clinical conditions. The MR‐EP catheters provided uncompromised recording of electrograms and cardiac pacing in combination with a standard EP recorder in MR‐guided in vivo EP studies, and the tip coil enabled fast and robust catheter localization. In vivo temperature measurements during such a study did not detect any device‐related heating, which confirms the high level of safety of the catheter, whereas unacceptable heating was found with a standard EP catheter. The presented concept for the first time enables catheters with full diagnostic EP functionality and active tracking and at the same time a sufficient level of radiofrequency safety for MRI without specific absorption rate‐related limitations. Magn Reson Med, 2011. © 2010 Wiley‐Liss, Inc.