An optical tomographic diagnostic instrument has been fielded at the Dual-Axis Radiographic Hydrodynamic Test Facility at Los Alamos National Laboratory. Four optical lines of sight create projections of an image of an electron beam on a Cerenkov target, which are relayed via optical fiber to streak cameras. From these projections, a reconstruction algorithm creates time histories of the beam's cross section. The instrument was fielded during and after facility commissioning, and tomographic reconstructions reported beam parameters. Results from reconstructions and analysis are noted.
An optical tomographic diagnostic instrument developed for the acquisition of high-speed time-resolved images has been fielded at the Dual-Axis Radiographic Hydrodynamic Test (DARHT) Facility at Los Alamos National Laboratory. The instrument was developed for the creation of time histories of electron-beam cross section through the collection of Cerenkov light. Four optical lines of sight optically collapse an image and relay projections via an optical fiber relay to recording instruments; a tomographic reconstruction algorithm creates the time history. Because the instrument may be operated in an adverse environment, it may be operated, adjusted, and calibrated remotely. The instrument was operated over the course of various activities during and after DARHT commissioning, and tomographic reconstructions reported verifiable beam characteristics. Results from the collected data and reconstructions and analysis of the data are discussed.
An instrument has been developed to acquire time-resolved tomographic data from the electron beam at the DARHT [Dual-Axis Radiographic Hydrodynamic Test] facility at Los Alamos National Laboratory. The instrument contains four optical lines of sight that view a single tilted object. The lens design optically integrates along one optical axis for each line of sight. These images are relayed via fiber optic arrays to streak cameras, and the recorded streaks are used to reconstruct the original two-dimensional data. Installation of this instrument into the facility requires automation of both the optomechanical adjustments and calibration of the instrument in a constrained space. Additional design considerations include compound tilts on the object and image planes.