Background: Vascular reconstructions are becoming challenging due to the comorbidity of the aging population and since the introduction of minimally invasive approaches. Many sutureless anastomosis devices have been designed to facilitate the cardiovascular surgeon's work and the vascular join (VJ) is one of these. We designed an animal study to assess its reliability and long-term efficacy. Methods: VJ allows the construction of end-to-end and end-to-side anastomoses. It consists of two metallic crowns fixed to the extremity of the two conduits so that vessel edges are joined layer by layer. There is no foreign material exposed to blood. In adult sheep both carotid arteries were prepared and severed. End-to-end anastomoses were performed using the VJ device on one side and the classical running suture technique on the other side. Animals were followed-up with Duplex-scan every 3 months and sacrificed after 12 months. Histopathological. analysis was carried out. Results: In 20 animals all 22 sutureless anastomoses were successfully completed in less than 2 min versus 6 +/- 3 min for running suture. Duplex showed the occlusion of three controls and one sutureless anastomosis. Two controls and one sutureless had stenosis > 50%. Histology showed very thin layer of myointimal hyperptasia (50 +/- 10 mu m) in the sutureless group versus 300 +/- 27 mu m in the control. No significant inflammatory reaction was detected. Conclusions: VJ provides edge-to-edge vascular repair that can be considered the most physiological way to restore vessel continuity. For the first time, in healthy sheep, an anastomotic device provided better results than suture technique. (c) 2006 Elsevier B.V. All rights reserved.
Five years of experience with endovascular infrarenal aneurysm repair at our institution is reviewed. Implantation of endoprostheses in 88 patients has been performed by surgeons using exclusively intravascular ultrasound (IVUS) and fluoroscopy. IVUS identified the target site of deployment in all cases. In-hospital morbidity was 22% (19/88). Two percent mortality (2/88) and 5% early conversion (4/88) as a consequence of type I endoleaks were noted only in the first 53 patients with early devices (NS). Early endoleaks were present in 36% (32/88) including twenty-two type I, five type II and five type III endoleaks. Proximal endoleaks were associated with early devices (P<0.001), and technical difficulties with deployment. Tube grafts used in the beginning, performed poorly with 54% (7/13) type I endoleaks. Endoleaks diminished to 10% (9/88) by spontaneous closure and secondary endovascular procedures that were necessary in 24% (21/88) and consisted of coil embolization/cuff extension (9), late conversion (6), and limb recanalization or femoral cross-over bypass (6). Endovascular aneurysm repair using IVUS is a valid alternative technique. Improved devices and systematic use of bifurcated endoprostheses for infrarenal aneurysms reduce the occurrence of type I endoleaks.