The scientific performance of the Planck Low Frequency Instrument (LFI) after one year of in-orbit operation is presented. We describe the main optical parameters and discuss photometric calibration, white noise sensitivity, and noise properties. A preliminary evaluation of the impact of the main systematic effects is presented. For each of the performance parameters, we outline the methods used to obtain them from the flight data and provide a comparison with pre-launch ground assessments, which are essentially confirmed in flight.
We present the calibration and scientific performance parameters of the Planck Low Frequency Instrument (LFI) measured during the ground cryogenic test campaign. These parameters characterise the instrument response and constitute our optimal pre-launch knowledge of the LFI scientific performance. The LFI shows excellent 1/f stability and rejection of instrumental systematic effects; its measured noise performance shows that LFI is the most sensitive instrument of its kind. The calibration parameters will be updated during flight operations until the end of the mission.
The Low Frequency Instrument (LFI) on-board the ESA Planck satellite carries eleven radiometer subsystems, called radiometer chain assemblies (RCAs), each composed of a pair of pseudo-correlation receivers. We describe the on-ground calibration campaign performed to qualify the flight model RCAs and to measure their pre-launch performances. Each RCA was calibrated in a dedicated flight-like cryogenic environment with the radiometer front-end cooled to 20 K and the back-end at 300 K, and with an external input load cooled to 4 K. A matched load simulating a blackbody at different temperatures was placed in front of the sky horn to derive basic radiometer properties such as noise temperature, gain, and noise performance, e. g. 1/f noise. The spectral response of each detector was measured as was their susceptibility to thermal variation. All eleven LFI RCAs were calibrated. Instrumental parameters measured in these tests, such as noise temperature, bandwidth, radiometer isolation, and linearity, provide essential inputs to the Planck-LFI data analysis.
In this paper we present the Low Frequency Instrument (LFI), designed and developed as part of the Planck space mission, the ESA program dedicated to precision imaging of the cosmic microwave background (CMB). Planck-LFI will observe the full sky in intensity and polarisation in three frequency bands centred at 30, 44 and 70 GHz, while higher frequencies (100-850 GHz) will be covered by the HFI instrument. The LFI is an array of microwave radiometers based on state-of-the-art Indium Phosphide cryogenic HEMT amplifiers implemented in a differential system using blackbody loads as reference signals. The front-end is cooled to 20K for optimal sensitivity and the reference loads are cooled to 4K to minimise low frequency noise. We provide an overview of the LFI, discuss the leading scientific requirements and describe the design solutions adopted for the various hardware subsystems. The main drivers of the radiometric, optical and thermal design are discussed, including the stringent requirements on sensitivity, stability, and rejection of systematic effects. Further details on the key instrument units and the results of ground calibration are provided in a set of companion papers.
The Planck Low Frequency Instrument (LFI) radiometers have been tested extensively during several dedicated campaigns. The present paper reports the principal noise properties of the LFI radiometers. A brief description of the LFI radiometers is given along with details of the test campaigns relevant to determination of noise properties. Current estimates of flight sensitivities, 1/f parameters, and noise effective bandwidths are presented. The LFI receivers exhibit exceptional 1/f noise, and their white noise performance is sufficient for the science goals of Planck.
The Low Frequency Instrument on board the PLANCK satellite is designed to give the most accurate map ever of the CMB anisotropy of the whole sky over a broad frequency band spanning 27 to 77 GHz. It is made of an array of 22 pseudo-correlation radiometers, composed of 11 actively cooled (20 K) Front End Modules (FEMs), and 11 Back End Modules (BEMs) at 300K, each FEM and BEM set comprising two radiometers. The connection between the two parts is made with rectangular Wave Guides (WGs). Considerations of very different nature (thermal, electromagnetic and mechanical), imposed stringent requirements on the WGs characteristics and drove their design. From the thermal point of view, the WG should guarantee good insulation between the FEM and the BEM sections to avoid overloading the cryocooler. On the other hand it is essential that the signals do not undergo excessive attenuation through the WG. Finally, given the different positions of the FEM modules behind the focal surface and the mechanical constraints given by the surrounding structures, different mechanical designs were necessary. A composite configuration of Stainless Steel and Copper was selected to satisfy all the requirements described. Given the complex shape and the considerable length (about 1.5-2 m) of the LFI WGs, manufacturing and testing the WGs was a challenge. 1Corresponding author. 2Present address: R&D Technology Development, Numoyx Italy s.r.l., via C. Olivetti 2, 20041 Agrate Brianza (Mi),
In this paper we present the test results of the qualification model (QM) of the LFI instrument, which is being developed as part of the ESA Planck satellite. In particular we discuss the calibration plan which has defined the main requirements of the radiometric tests and of the experimental setups. Then we describe how these requirements have been implemented in the custom-developed cryo-facilities and present the main results. We conclude with a discussion of the lessons learned for the testing of the LFI Flight Model (FM).
Jodrell Bank Observatory is responsible for the Front end modules (FEMs) for 30 GHz for LFI, with IAC being responsible for the hybrids and any phase switch. The University of Cantabria are responsible for the Back end modules (BEMs). Developments have been made to test prototypes both warm and cold for the FEMs. This is evolving towards flight hardware through an elegant breadboard phase EBB. Results of system performance show both white noise and 1/f characteristics.
The description of the design, implementation and test results of a differential radiometer at 30 GAz with a 20 per cent of effective bandwidth for the Planck mission is presented. This Elegant Breadboard (EBB) 30 GHz radiometer is composed of a Front-End Module (FEM) assembled at Jodrell Bank Observatory (UK) and a Back- End Module (BEM) assembled at the University of Cantabria and Technical University of Catalonia (Spain). The achieved system noise temperature was lower than 10 Kelvin using state-of-the-art InP HEMT devices. Several tests have been carried out with the FEM cooled at cryogenic temperature (20 K) and the BEM at room temperature demonstrating system noise temperature, gain, effective bandwidth and 1/f noise cancellation according to its requirements.
The LFI (Low Frequency Instrument) on board the ESA Planck satellite is constituted by an array of radiometric detectors actively cooled at 20 K in the 30-70 GHz frequency range in the focal plane of the Planck telescope. In this paper we present an overview of the LFI instrument, with a particular focus on the radiometer design. The adopted pseudo-correlation scheme uses a software balancing technique (with a tunable parameter called gain modulation factor) which is effective in reducing the radiometer susceptibility to amplifier instabilities also in presence of small non-idealities in the radiometric chain components, provided that the gain modulation factor is estimated with an accuracy of the order of 0.2%. These results have been recently confirmed by experimental laboratory measurements conducted on the LFI prototype radiometers at 30, 70 and 100 GHz.