A fundamental parameter to evaluate the beam delivery precision and stability on a clinical linear accelerator (linac) is the focal spot position (FSP) measured relative to the collimator axis of the radiation head. The aims of this work were to evaluate comprehensive data on FSP acquired on linacs in clinical use and to establish the ability of alternative phantoms to detect effects on patient plan delivery related to FSP. FSP measurements were conducted using a rigid phantom holding two ball-bearings at two different distances from the radiation source. Images of these ball-bearings were acquired using the electronic portal imaging device (EPID) integrated with each linac. Machine QA was assessed using a radiation head-mounted PTW STARCHECK phantom. Patient plan QA was investigated using the SNC ArcCHECK phantom positioned on the treatment couch, irradiated with VMAT plans across a complete 360° gantry rotation and three X-ray energies. This study covered eight Elekta linacs, including those with 6 MV, 18 MV, and 6 MV flattening-filter-free (FFF) beams. The largest range in the FSP was found for 6 MV FFF. The FSP of one linac, retrofitted with 6 MV FFF, displayed substantial differences in FSP compared to 6 MV FFF beams on other linacs, which all had FSP ranges less than 0.50 mm and 0.25 mm in the lateral and longitudinal directions, respectively. The PTW STARCHECK phantom proved effective in characterising the FSP, while the SNC ArcCHECK measurements could not discern FSP-related features. Minor variations in FSP may be attributed to adjustments in linac parameters, component replacements necessary for beam delivery, and the wear and tear of various linac components, including the magnetron and gun filament. Consideration should be given to the ability of any particular phantom to detect a subsequent impact on the accuracy of patient plan delivery.
PurposeIonisation chamber based reference dosimetry in magnetic resonance linear accelerators (MRL) aimed for radiotherapy requires correction for recombination losses. Published studies have found that such corrections can be carried out using the two-voltage method. These studies have, however, not included comparison with recombination corrections based on the Niatel method, which can be seen as a robust reference method due to its clear separation of initial and volume recombination and its explicit account of the pulsed nature of the dose delivery. The primary objective of this work therefore was to carry out such a comparison.Materials and methodsFour Farmer-type chambers (PTW-30006 and PTW-30013) were placed in a water phantom in 1.5 T Elekta Unity MRL. The chambers were oriented antiparallel or perpendicular to the static magnetic field B0 and irradiated at a source-to-surface distance of 133.5 cm with a 10 × 10 cm2 field size.ResultsThe two-voltage method gave results in agreement (within 0.1%) with the recombination corrections derived from the Niatel method. The recombination corrections from three Niatel parameter sets (one based on a Varian Truebeam and two obtained directly in the MRL) deviated less than 0.1% from each other. A systematic shift in the recombination correction of less than 0.05% was observed if polarity corrections were not applied.ConclusionsThe study supports the use of the two-voltage method in MRLs based on its excellent agreement with the Niatel method. This work, therefore, complements existing knowledge as previous studies have not included a comparison with the Niatel method.
Current international protocols such as the IAEA/AAPM TRS-483 for small field dosimetry do not include recommendations for measurements in strong magnetic fields as found in MR-Linacs (i.e. megavoltage x-ray radiotherapy units with built-in magnetic resonance imaging). A EURAMET research project was therefore launched to support the traceability of such measurements and to contribute to the metrological basis for future recommendations. Given the high degree of water equivalence and the small size of fiber-coupled plastic scintillators, one objective of the project was to study the suitability of this detector type for dosimetry in MR-Linacs and to demonstrate their use in a clinical environment. This manuscript reports results of that work. A protocol for measuring field output factors with plastic scintillation detectors was developed. The protocol is based on calibration of the scintillators directly in a water phantom positioned in the MR-Linac rather than in a water-equivalent solid phantom as has been a common technique for conventional linacs. A detailed uncertainty budget was established involving all known sources of uncertainty including the systematic influence of light attenuation through the fiber when measuring in different field sizes. The protocol was used to measure field output factors at five MR-Linac’s (four Viewray MRIdian’s and one Elekta Unity). The protocol showed a high level of robustness leading to a less than 0.3 % deviation between field output factors calculated based on calibration data obtained before or after the field-output factor measurements. The correction for light attenuation was less than 0.4 % for all fields measured but showed a systematic offset, with larger corrections needed for small field sizes.This robustness of the protocol was further demonstrated by the closeness of agreement (agreement within uncertainty) for field-output factor measurements with two different scintillators (BCF-12 and BCF-60) for square fields (size: 0.83 cm x 0.83 cm or larger) in four nominally identical Viewray MRIdian's. To test the accuracy of the scintillator measurements, the system was used for characterization of a PTW 60019 micro diamond detector in a Viewray MRIdian. Scintillator based correction factors for output factor measurements with the micro diamond detector were found to be in agreement with independent Monte-Carlo computations. The study therefore supports that plastic scintillation detectors are an excellent option for measuring field output factors in MR-Linacs.
PURPOSE:The precision of the dose delivery in radiation therapy with high-field MR-linacs is challenging due to the substantial variation in the beam attenuation of the patient positioning system (PPS) (the couch and coils) as a function of the gantry angle. This work aimed to compare the attenuation of two PPSs located at two different MR-linac sites through measurements and calculations in the treatment planning system (TPS).METHODS:Attenuation measurements were performed at every 1° gantry angle at the two sites with a cylindrical water phantom with a Farmer chamber inserted along the rotational axis of the phantom. The phantom was positioned with the chamber reference point (CRP) at the MR-linac isocentre. A compensation strategy was applied to minimise sinusoidal measurement errors due to, e.g. air cavity or setup. A series of tests were performed to assess the sensitivity to measurement uncertainties. The dose to a model of the cylindrical water phantom with the PPS added was calculated in the TPS (Monaco v5.4 as well as in a development version Dev of an upcoming release), for the same gantry angles as for the measurements. The TPS PPS model dependency of the dose calculation voxelisation resolution was also investigated.RESULTS:A comparison of the measured attenuation of the two PPSs yielded differences of less than 0.5% for most gantry angles. The maximum deviation between the attenuation measurements for the two different PPSs exceeded ±1% at two specific gantry angles 115° and 245°, where the beam traverses the most complex PPS structures. The attenuation increases from 0% to 25% in 15° intervals around these angles. The measured and calculated attenuation, as calculated in v5.4, was generally within 1-2% with a systematic overestimation of the attenuation for gantry angles around 180°, as well as a maximum error of 4-5% for a few discrete angles in 10° gantry angle intervals around the complex PPS structures. The PPS modelling was improved compared to v5.4 in Dev, especially around 180°, and the results of those calculations were within ±1%, but with a similar 4% maximum deviation for the most complex PPS structures.CONCLUSIONS:Generally, the two tested PPS structures exhibit very similar attenuation as a function of the gantry angle, including the angles with a steep change in attenuation. Both TPS versions, v5.4 and Dev delivered clinically acceptable accuracy of the calculated dose, as the differences in the measurements were overall better than ±2%. Additionally, Dev improved the accuracy of the dose calculation to ±1% for gantry angles around 180°.
The magnetic field in magnetic resonance imaging-guided radiotherapy (MRgRT) systems influences the three-dimensional (3D) dose deposition and hence the delivered dose distributions. The aim of this study was to investigate the dose-rate dependency and dose response of silicone-based radiochromic dosimeters for photon irradiation in the precense of a magnetic field using a 1.5 T MRgRT system. Additionally, the study aimed to provide a proof of the concept of radiotherapy treatment based on a treatment plan calculated on magnetic resonance imaging rather than a computed tomography (CT) scan. The delivered dose was read out in 3D with an optical CT scanner and the dose distribution was verified using gamma analysis. We found an insignificant dose-rate dependency for dose rates ranging from 3.2 to 5.1 Gy/min and a linear dose response up to 20 Gy. A 3D 3%/3mm gamma analysis showed a pass rate of 95.1%. The dosimeter showed clinical potential for 3D dose verification of MRgRT delivery.
The recent introduction of a commercial 1.5 T MR-linac system has considerably improved the image quality of the patient acquired in the treatment unit as well as enabling online adaptive radiation therapy (oART) treatment strategies. Quality Assurance (QA) of this new technology requires new methodology that allows for the high field MR in a linac environment. The presence of the magnetic field requires special attention to the phantoms, detectors, and tools to perform QA. Due to the design of the system, the integrated megavoltage imager (MVI) is essential for radiation beam calibrations and QA. Additionally, the alignment between the MR image system and the radiation isocenter must be checked. The MR-linac system has vendor-supplied phantoms for calibration and QA tests. However, users have developed their own routine QA systems to independently check that the machine is performing as required, as to ensure we are able to deliver the intended dose with sufficient certainty. The aim of this work is therefore to review the MR-linac specific QA procedures reported in the literature.
Objective. The objective of this study was to separately quantify the stability of the megavoltage imager (MVI) and radiation head of an Elekta Unity MRL, throughout full gantry rotation.Approach. A ball-bearing (BB) phantom was attached to the radiation head of the Unity, while a single BB was placed at isocentre. Images were acquired during rotation, using the MVI. These images were processed using an in-house developed MATLAB program to reduce the errors resulted by noise, and the positions of the BBs in the images were analysed to extract MVI and radiation head sag data.Main results. The results returned by this method showed reproducibility, with a mean standard deviation of 7μm for the position of BBs across all gantry angles. The radiation head was found to sag throughout rotation, with a maximum course of movement of 0.59 mm. The sag pattern was stable over a period greater than a year but showed some dependence on gantry rotation direction.Significance. As MRL is a relatively new system, it is promising to have data supporting the high level of precision on one Elekta Unity machine. Isolating and quantifying the sources of uncertainty in radiation delivery may allow more sophisticated analysis of how the system performance may be improved.
The clinical introduction of hybrid magnetic resonance (MR) guided radiotherapy (RT) delivery systems has led to the need to validate the end-to-end dose delivery performance on such machines. In the current study, an MR visible phantom was developed and used to test the spatial deviation between planned and delivered dose at two 1.5 T MR linear accelerator (MR linac) systems, including pre-treatment imaging, dose planning, online imaging, image registration, plan adaptation, and dose delivery. The phantom consisted of 3D printed plastic and MR visible silicone rubber. It was designed to minimise air gaps close to the radiochromic film used as a dosimeter. Furthermore, the phantom was designed to allow submillimetre, reproducible positioning of the film in the phantom. At both MR linac systems, 54 complete adaptive, MR guided RT workflow sessions were performed. To test the dose delivery performance of the MR linac systems in various adaptive RT (ART) scenarios, the sessions comprised a range of systematic positional shifts of the phantom and imaging or plan adaptation conditions. In each workflow session, the positional translation between the film and the adaptive planned dose was determined. The results showed that the accuracy of the MR linac systems was between 0.1 and 0.9 mm depending on direction. The highest mean deviance observed was in the posterior-anterior direction, and the direction of the error was consistent between centres. The precision of the systems was related to whether the workflow utilized the internal image registration algorithm of the MR linac. Workflows using the internal registration algorithm led to a worse precision (0.2-0.7 mm) compared to workflows where the algorithm was decoupled (0.2 mm). In summary, the spatial deviation between planned and delivered dose of MR-guided ART at the two MR linac systems was well below 1 mm and thus acceptable for clinical use.
The potential for delivering high precision radiotherapy using linear accelerators (linacs) has been improved with the development of digital x-ray electronic portal imaging devices (EPID) for acquiring kilovoltage (kV) cone-beam CT and megavoltage (MV) images for patient positioning. EPIDs have also opened the possibilities of developing novel quality assurance and insight into radiotherapy equipment performance. The aim of this work was to measure the offset of the focal spot position (FSP) of a linac under gantry rotation relative to the collimator axis using an EPID. The focal spot was assumed to be a point source of MV x-ray generation. A special phantom was designed for measurement of FSP as a function of gantry angle on clinical linacs. The phantom was designed for attachment to the gantry head and supporting two tungsten-carbide ball-bearings at two different distances from the focal spot. The methodology was demonstrated on a series of images acquired of the phantom on three Elekta linacs in clinical use with 6 MV flattening-filter-free (FFF) beams. The gantry and collimator were rotated 360° in steps of 30°. For each position an image of the phantom was acquired using the EPID. Each series consisted of 169 EPID images. The images were analysed using in-house developed software. Analyses of the EPID images acquired with 6 MV FFF beams showed that the focal spot motion amplitudes relative to the collimator axis during gantry rotation in the longitudinal and lateral directions were less than 0.10 mm and 0.50 mm, respectively, for an optimized 6 MV FFF FSP calibrated linac. In a treatment planning system (TPS) the focal spot is assumed to be located on the rotation axis of the collimator at all gantry angles. This work introduces a method for quantifying the actual variation from this assumption in practice.
Purpose: A 1.5 T MR Linac (MRL) has recently become available. MRL treatment workflows (WF) include online plan adaptation based on daily MR images (MRI). This study reports initial clinical experiences after five months of use in terms of patient compliance, cases, WF timings, and dosimetric accuracy. Method and materials: Two different WF were used dependent on the clinical situation of the day; Adapt To Position WF (ATP) where the reference plan position is adjusted rigidly to match the position of the targets and the OARs, and Adapt To Shape WF (ATS), where a new plan is created to match the anatomy of the day, using deformable image registration. Both WFs included three 3D MRI scans for plan adaptation, verification before beam on, and validation during IMRT delivery. Patient compliance and WF timings were recorded. Accuracy in dose delivery was assessed using a cylindrical diode phantom. Results: Nineteen patients have completed their treatment receiving a total of 176 fractions. Cases vary from prostate treatments (60Gy/20F) to SBRT treatments of lymph nodes (45 Gy/3F) and castration by ovarian irradiation (15 Gy/3F). The median session time (patient in to patient out) for 127 ATPs was 26 (21-78) min, four fractions lasted more than 45 min due to additional plan adaptation. For the 49 ATSs a median time of 12 (1-24) min was used for contouring resulting in a total median session time of 42 (29-91) min. Three SBRT fractions lasted more than an hour. The time on the MRL couch was well tolerated by the patients. The median gamma pass rate (2 mm,2% global max) for the adapted plans was 99.2 (93.4-100)%, showing good agreement between planned and delivered dose. Conclusion: MRL treatments, including daily MRIs, plan adaptation, and accurate dose delivery, are possible within a clinically acceptable timeframe and well tolerated by the patients.
ConclusionUsing 3D printed MR visible silicone inserts it was possible to create an end-to-end validation phantom for MR linacs.The phantom has been used in our department and the uncertainty of the dose positions is around half a millimeter, which is needed to make precise MR linac treatments.