Background Conventional chest and abdominal MRI require breath-holds to reduce motion artifacts. Neonates and infants require general anesthesia with intubation to enable breath-held acquisitions. Objective We aimed to validate a free-breathing approach to reduce general anesthesia using a motion-insensitive radial acquisition with respiratory gating. Materials and methods We retrospectively enrolled children <3 years old who were referred for MRI of the chest or abdomen. They were divided into two groups according to MRI protocol: (1) breath-held scans under general anesthesia with T2-weighted single-shot fast spin-echo (SSFSE) and contrast-enhanced T1-weighted modified Dixon, and (2) free-breathing scans using radial sequences (T2-W MultiVane XD and contrast-enhanced T1-W three-dimensional [3-D] Vane XD). Two readers graded image quality and motion artifacts. Results We included 23 studies in the free-breathing cohort and 22 in the breath-hold cohort. The overall imaging scores for the free-breathing radial T2-W sequence were similar to the scores for the breath-held T2-W SSFSE sequence (chest, 3.6 vs. 3.2, P=0.07; abdomen, 3.9 vs. 3.7, P=0.66). The free-breathing 3-D radial T1-W sequence also had image quality scores that were similar to the breath-held T1-W sequence (chest, 4.0 vs. 3.0, P=0.06; abdomen, 3.7 vs. 3.9, P=0.15). Increased motion was seen in the abdomen on the radial T2-W sequence (P<0.001), but increased motion was not different in the chest (P=0.73) or in contrast-enhanced T1-W sequences (chest, P=0.39; abdomen, P=0.15). The mean total sequence time was longer in free-breathing compared to breath-held exams (P<0.01); however, this did not translate to longer overall exam times (P=0.94). Conclusion Motion-insensitive radial sequences used for infants and neonates were of similar image quality to breath-held sequences and had decreased sedation and intubation
Conventional pediatric volumetric MRI acquisitions of a short-axis stack typically require multiple breath-holds under anesthesia. Here, we aimed to validate a vendor-optimized compressed-sensing approach to reduce scan time during short-axis balanced steady-state free precession (bSSFP) cine imaging. Imaging was performed in 28 patients (16±9 years) in this study on a commercial 3-tesla (T) scanner using retrospective electrocardiogram-gated cine bSSFP. Cine short-axis images covering both ventricles were acquired with conventional parallel imaging and a vendor-optimized parallel imaging/compressed-sensing approach. Qualitative Likert scoring for blood–myocardial contrast, edge definition, and presence of artifact was performed by two experienced radiologists. Quantitative comparisons were performed including biventricular size and function. A paired t-test was used to detect significant differences (P<0.05). Scan duration was 7±2 s/slice for conventional imaging (147±33 s total) vs. 4±2 s/slice for compressed sensing (83±28 s total). No significant differences were found with qualitative image scores for blood–myocardial contrast, edge definition, and presence of artifact. No significant differences were found in volumetric analysis between the two sequences. The number of breath-holds was 10±4 for conventional imaging and 5±3 for compressed sensing. Compressed sensing allowed for a 50% reduction in the number of breath-holds and a 43% reduction in the total scan time without differences in the qualitative or quantitative measurements as compared to the conventional technique.