Background: High-intensity focused ultrasound (HIFU) has been shown to control bleeding from liver incisions, and blood vessel punctures and incisions. The objective of the current study was to investigate the capability of HIFU to stop bleeding from splenic injuries in a pig model. Methods: Surgical incisions, 25 to 50 mm in length and 2 to 8 mm in depth, were made in the spleens of five anesthetized pigs. HIFU with a frequency of 5 MHz was applied within 5 seconds of making the incision. A total of 39 incisions and HIFU treatments were performed. Results: Bleeding from all incisions was stopped completely after HIFU treatment. The average times to control and completely arrest the hemorrhage were 28 and 55 seconds, respectively. The mechanisms of hemostasis appeared to be thermally induced coagulation necrosis of splenic tissue and occlusion of blood vessels by a mechanically induced homogenized splenic tissue. Conclusion: HIFU may provide a useful method of hemostasis for actively bleeding spleen. Because of its ability to induce hemostasis at adjustable depth, HIFU may prove to be a useful cauterization method both in the operating room and for patients who are managed nonoperatively.
—The use of Doppler ultrasound was investigated to determine if it would aid in guiding the application of high-intensity focused ultrasound (HIFU) to stop bleeding from punctured vessels. Major vessels (abdominal aorta, illiac, carotid, common femoral and superficial femoral arteries and the jugular vein) were surgically exposed, punctured and treated in anesthetized pigs. Treatment was applied when the Doppler sounds indicated the focus coincided with the bleeding site. In 89 treatment trials, the average time to achieve major hemostasis (a point where bleeding was reduced to a level of only oozing) was 8 s, and for complete hemostasis was 13 s. These times were significantly shorter than those of an identical former study in which only visual guidance was used. In that study, the average times for major and complete hemostasis were 40 and 62 s, respectively. The advantage of Doppler guidance in applying HIFU in treating bleeding vessels was demonstrated.
Objective: High-intensity focused ultrasound (HIFU) has been shown to be effective in controlling hemorrhage from punctures in blood vessels. The objective of the current study was to investigate the capability of HIFU to stop bleeding after a more severe type of vascular injury, namely longitudinal incisions of arteries and veins.Methods: The superficial femoral arteries, common femoral arteries, carotid arteries, and jugular veins of four anesthetized pigs were exposed surgically. A longitudinal incision, 2 to 8 mm in length, was produced in the vessel. HIFU treatment was applied within 5 seconds of the onset of the bleeding. The HIFU probe consisted of a high-power, 3.5-MHz, piezoelectric transducer with an ellipsoidal focal spot that was 1 mm in cross section and 9 mm in axial dimension. The entire incision area was scanned with the HIFU beam at a rate of 15 to 25 times/second and a linear displacement of 5 to 10 mm. A total of 76 incisions and HIFU treatments were performed.Results: Control of bleeding (major hemosatsis) was achieved in all 76 treatments, with complete hemostasis achieved in 69 treatments (91%). The average treatment times of major and complete hemostasis were 17 and 25 seconds, respectively. After the treatment, 74% of the vessels in which complete hemostasis was achieved were patent with distal blood flow and 26% were occluded. The HIFU-treated vessels showed a consistent coagulation of the adventitia surrounding the vessels, with a remarkably localized injury to the vessel wall. Extensive fibrin deposition at the treatment site was observed.Conclusion: HIFU may provide a useful method of achieving hemostasis for arteries and veins in a variety of clinical applications.
The hemorrhagic complications of vascular injury can be significant. We report on the use of high-intensity focused ultrasound (HIFU) to stop the hemorrhage of punctured blood vessels in pigs. Two HIFU transducers with frequencies of 3.5 and 2.0 MHz, each equipped with a water-filled conical housing, were used. Major blood vessels (femoral artery and vein, axillary artery, carotid artery and jugular vein), 2-10 mm in diameter, of anesthetized pigs were exposed surgically and punctured with 14- and 18-gauge needles to produce moderate to profuse bleeding. Complete hemostasis was achieved in less than 3 min of HIFU treatment in most blood vessels, and all vessels were patent after the treatment. Both HIFU frequencies were effective in producing hemostasis. Gross examination of the HIFU-treated vessels showed a consistent hardening of the soft tissue surrounding the blood vessels, providing a seal for the puncture hole. Microscopic examination of the vessels showed a remarkably localized HIFU treatment, resulting in coagulation of the adventitia, and an extensive fibrin network around the vessels and in the puncture hole. The vessel walls exhibited focal swelling, without evidence of irreversible injury. HIFU may provide a useful method for achieving hemostasis of punctured and traumatized blood vessels in a variety of clinical settings.
A severe arterial occlusion in the leg usually is bypassed by implanting a saphenous vein harvested from the limb. Once implanted, the vein functions well but over time may develop stenoses that may lead to occlusion. In order to detect and correct the stenoses that may lead to graft failure, frequent surveillance of the vein graft is required. A new ultrasound imaging method was developed to display the panoramic view of the vein graft in combination with its blood flow velocity waveform for surveillance. The panoramic view is the projection (ray-casting) image of multiple B-mode images with sequential longitudinal view of the vein graft. The velocity waveform also is recorded along the vessel with pulsed Doppler ultrasound. The acquired images and waveforms from the ultrasound scanner are registered individually in three-dimensional space with an electromagnet-based position and orientation sensor located on the scanhead. A prominent point on the scar from the surgery is used as the fiducial mark for spatial registration, so that the same lesion in the vein graft can be tracked automatically at each visit for retrospective study.