Abstract Introduction Commissioning, calibration, and quality control procedures for nuclear medicine imaging systems are typically performed using hollow containers filled with radionuclide solutions. This leads to multiple sources of uncertainty, many of which can be overcome by using traceable, sealed, long-lived surrogate sources containing a radionuclide of comparable energies and emission probabilities. This study presents the results of a quantitative SPECT/CT imaging comparison exercise performed within the MRTDosimetry consortium to assess the feasibility of using 133Ba as a surrogate for 131I imaging. Materials and methods Two sets of four traceable 133Ba sources were produced at two National Metrology Institutes and encapsulated in 3D-printed cylinders (volume range 1.68–107.4 mL). Corresponding hollow cylinders to be filled with liquid 131I and a mounting baseplate for repeatable positioning within a Jaszczak phantom were also produced. A quantitative SPECT/CT imaging comparison exercise was conducted between seven members of the consortium (eight SPECT/CT systems from two major vendors) based on a standardised protocol. Each site had to perform three measurements with the two sets of 133Ba sources and liquid 131I. Results As anticipated, the 131I pseudo-image calibration factors (cps/MBq) were higher than those for 133Ba for all reconstructions and systems. A site-specific cross-calibration reduced the performance differences between both radionuclides with respect to a cross-calibration based on the ratio of emission probabilities from a median of 12–1.5%. The site-specific cross-calibration method also showed agreement between 133Ba and 131I for all cylinder volumes, which highlights the potential use of 133Ba sources to calculate recovery coefficients for partial volume correction. Conclusion This comparison exercise demonstrated that traceable solid 133Ba sources can be used as surrogate for liquid 131I imaging. The use of solid surrogate sources could solve the radiation protection problem inherent in the preparation of phantoms with 131I liquid activity solutions as well as reduce the measurement uncertainties in the activity. This is particularly relevant for stability measurements, which have to be carried out at regular intervals.
Validation of a Molecular Radiotherapy (MRT) dosimetry system requires imaging data for which an accompanying "ground truth" pharmacokinetic model and absorbed dose calculation are known.Methods:We present a methodology for production of a validation dataset for image based 177Lu dotatate dosimetry calculations. A pharmacokinetic model is presented with activity concentrations corresponding to common imaging timepoints. Anthropomorphic 3D printed phantoms, corresponding to the organs at risk, have been developed to provide SPECT/CT and Whole Body imaging with known organ activities corresponding to common clinical timepoints.Results:Results for the accuracy of phantom filling reproduce the activity concentrations from the pharmacokinetic model for all timepoints and organs within measurement uncertainties, with a mean deviation of 0.6(8)%. The imaging dataset, ancillary data and phantoms designs are provided as a source of well characterized input data for the validation of clinical MRT dosimetry systems.Conclusions:The combination of pharmacokinetic modelling with the use of anthropomorphic 3D printed phantoms are a promising procedure to provide data for the validation of Molecular Radiotherapy Dosimetry systems, allowing multicentre comparisons.
PURPOSE:Monte Carlo modelling of SPECT imaging in Molecular Radiotherapy can improve activity quantification. Until now, SPECT modelling with GATE only considered circular orbit (CO) acquisitions. This cannot reproduce auto-contour acquisitions, where the detector head moves close to the patient to improve image resolution. The aim of this work is to develop and validate an auto-contouring step-and-shoot acquisition mode for GATE SPECT modelling. METHODS:177Lu and 131I SPECT experimental acquisitions performed on a Siemens Symbia T2 and GE Discovery 670 gamma camera, respectively, were modelled. SPECT projections were obtained for a cylindrical Jaszczak phantom and a lung and spine phantom. Detector head parameters (radial positions and acquisition angles) were extracted from the experimental projections to model the non-circular orbit (NCO) detector motion. The gamma camera model was validated against the experimental projections obtained with the cylindrical Jaszczak (177Lu) and lung and spine phantom (131I). Then, 177Lu and 131I CO and NCO SPECT projections were simulated to validate the impact of explicit NCO modelling on simulated projections. RESULTS:Experimental and simulated SPECT images were compared using the gamma index, and were in good agreement with gamma index passing rate (GIPR) and gammaavg of 96.27%, 0.242 (177Lu) and 92.89%, 0.36 (131I). Then, simulated 177Lu and 131I CO and NCO SPECT projections were compared. The GIPR, gammaavg between the two gamma camera motions was 99.85%, 0.108 for 177Lu and 75.58%, 0.6 for 131I. CONCLUSION:This work thereby justifies the need for auto-contouring modelling for isotopes with high septal penetration.
Purpose: Monte Carlo modelling of SPECT imaging in Molecular Radiotherapy can improve activity quantifica-tion. Until now, SPECT modelling with GATE only considered circular orbit (CO) acquisitions. This cannot reproduce auto-contour acquisitions, where the detector head moves close to the patient to improve image resolution. The aim of this work is to develop and validate an auto-contouring step-and-shoot acquisition mode for GATE SPECT modelling.& nbsp;Methods: Lu-177 and I-131 SPECT experimental acquisitions performed on a Siemens Symbia T2 and GE Discovery 670 gamma camera, respectively, were modelled. SPECT projections were obtained for a cylindrical Jaszczak phantom and a lung and spine phantom. Detector head parameters (radial positions and acquisition angles) were extracted from the experimental projections to model the non-circular orbit (NCO) detector motion. The gamma camera model was validated against the experimental projections obtained with the cylindrical Jaszczak (Lu-177) and lung and spine phantom (I-131). Then, Lu-177 and I-131 CO and NCO SPECT projections were simulated to validate the impact of explicit NCO modelling on simulated projections.& nbsp;Results: Experimental and simulated SPECT images were compared using the gamma index, and were in good agreement with gamma index passing rate (GIPR) and gammaavg of 96.27%, 0.242 (177Lu) and 92.89%, 0.36 (131I). Then, simulated Lu-177 and I-131 CO and NCO SPECT projections were compared. The GIPR, gammaavg between the two gamma camera motions was 99.85%, 0.108 for Lu-177 and 75.58%, 0.6 for I-131.& nbsp;Conclusion: This work thereby justifies the need for auto-contouring modelling for isotopes with high septal penetration.
Purpose Patient-specific dosimetry is required to ensure the safety of molecular radiotherapy and to predict response. Dosimetry involves several steps, the first of which is the determination of the activity of the radiopharmaceutical taken up by an organ/lesion over time. As uncertainties propagate along each of the subsequent steps (integration of the time–activity curve, absorbed dose calculation), establishing a reliable activity quantification is essential. The MRTDosimetry project was a European initiative to bring together expertise in metrology and nuclear medicine research, with one main goal of standardizing quantitative 177 Lu SPECT/CT imaging based on a calibration protocol developed and tested in a multicentre inter-comparison. This study presents the setup and results of this comparison exercise. Methods The inter-comparison included nine SPECT/CT systems. Each site performed a set of three measurements with the same setup (system, acquisition and reconstruction): (1) Determination of an image calibration for conversion from counts to activity concentration (large cylinder phantom), (2) determination of recovery coefficients for partial volume correction (IEC NEMA PET body phantom with sphere inserts), (3) validation of the established quantitative imaging setup using a 3D printed two-organ phantom (ICRP110-based kidney and spleen). In contrast to previous efforts, traceability of the activity measurement was required for each participant, and all participants were asked to calculate uncertainties for their SPECT-based activities. Results Similar combinations of imaging system and reconstruction lead to similar image calibration factors. The activity ratio results of the anthropomorphic phantom validation demonstrate significant harmonization of quantitative imaging performance between the sites with all sites falling within one standard deviation of the mean values for all inserts. Activity recovery was underestimated for total kidney, spleen, and kidney cortex, while it was overestimated for the medulla. Conclusion This international comparison exercise demonstrates that harmonization of quantitative SPECT/CT is feasible when following very specific instructions of a dedicated calibration protocol, as developed within the MRTDosimetry project. While quantitative imaging performance demonstrates significant harmonization, an over- and underestimation of the activity recovery highlights the limitations of any partial volume correction in the presence of spill-in and spill-out between two adjacent volumes of interests.
The motivation of this study derives from the need for tumour margin estimation after surgical excision. Conventional beta autoradiography of beta emitters can be used to image tissue sections providing high spatial resolution compared to in-vivo molecular imaging. However, it requires sectioning of the specimen and it provides a 2D image of the tissue. Imaging of the 3D tissue sample can be achieved either by imaging sequential 2D sections, which is time-consuming and laborious, or by using a specialised detector for imaging that records the particles’ direction, in addition to their position, when they hit the detector. In this work we investigate whether a novel beta-tomosynthesis approach can be used for depth resolution of beta emitters. The technique involves acquiring multiple 2D images of the intact tissue sample while the detector rotates around the sample. The images are then combined and used to reconstruct the 3D position of the sources from a limited angle of conventional 2D autoradiography images. We present the results from Geant4 forward simulations and the reconstructed images from a breast tissue sample containing a Fluorine-18 positron emission source. The experiments show that the proposed method can provide depth resolution under certain conditions, indicating that there is potential for its use as a 3D molecular imaging technique of surgical samples in the future.
The optimised delivery of Molecular Radiotherapy requires individualised calculation of absorbed dose to both targeted lesions and neighbouring healthy tissue. To achieve this, accurate quantification of the activity distribution in the patient by external detection is vital.MethodsThis work extends specific anatomy-related calibration to true organ shapes. A set of patient-specific 3D printed organ inserts based on a diagnostic CT scan was produced, comprising the liver, spleen and both kidneys. The inserts were used to calculate patient-specific calibration factors for 177Lu. These calibration factors were compared with previously reported calibration factors for corresponding organ models based on the Cristy and Eckerman phantom series and for a comparably sized sphere. Monte Carlo calculations of the patient-specific radiation dose were performed for comparison with current clinical dosimetry methods for these data.ResultsPatient-specific calibration factors are shown to be dependent on the volume, shape and position of the organ containing activity with a corresponding impact on the calculation of the dose to the patient. The impact of organ morphology on calculated dose is reduced when the dominant contributor to dose is beta particles. This is due to the small range of beta particles in tissue. Overestimations of recovered activity and hence dose of up to 135% are observed.ConclusionFor accurate quantification to be performed calibration factors accounting for organ size, shape and position must be used. Such quantification is vital if accurate, patient-specific dosimetry is to be achieved.
Accurate image quantification requires accurate calibration of the detector and is vital if dosimetry is to be performed in molecular radiotherapy. A dependence on the position of calibration has been observed in single photon emission computed tomography images when attenuation correction (AC) and scatter correction are applied. This work investigates the origin of this dependence in single photon emission computed tomography scans of phantom inserts filled with Lu solution. A 113 ml sphere and inserts representing a mathematical model of a spleen and an anatomical model of a patient spleen were imaged at the centre and edge of elliptical phantoms. For these inserts, the difference in calibration factor between the positions was around 10% for images reconstructed with AC and triple energy window scatter correction. A combination of experimental imaging and Monte Carlo simulation was used to isolate possible causes due to imaging or reconstruction in turn. Inconsistent application of AC between different reconstruction systems was identified as the origin of the positional dependence.
Breast cancer is the most common cancer in women in the world. Breast-conserving surgery (BCS) is a standard surgical treatment for breast cancer with the key objective of removing breast tissue, maintaining a negative surgical margin and providing a good cosmetic outcome. A positive surgical margin, meaning the presence of cancerous tissues on the surface of the breast specimen after surgery, is associated with local recurrence after therapy. In this study, we investigate a new imaging modality based on Cerenkov luminescence imaging (CLI) for the purpose of detecting positive surgical margins during BCS. We develop Monte Carlo (MC) simulations using the Geant4 nuclear physics simulation toolbox to study the spectrum of photons emitted given 18F-FDG and breast tissue properties. The resulting simulation spectra show that the CLI signal contains information that may be used to estimate whether the cancerous cells are at a depth of less than 1 mm or greater than 1 mm given appropriate imaging system design and sensitivity. The simulation spectra also show that when the source is located within 1 mm of the surface, the tissue parameters are not relevant to the model as the spectra do not vary significantly. At larger depths, however, the spectral information varies significantly with breast optical parameters, having implications for further studies and system design. While promising, further studies are needed to quantify the CLI response to more accurately incorporate tissue specific parameters and patient specific anatomical details.
BACKGROUND:Deprivation is associated with an increased risk of invasive Neisseria meningitidis disease, but little is known about the relationship between deprivation and asymptomatic carriage of N. meningitidis. This analysis was conducted to examine the relationship between meningococcal carriage and deprivation.METHODS:As part of a rapid meningococcal carriage prevalence study conducted in West Cumbria to investigate an apparent cluster of invasive meningococcal disease, data were collected on lifestyle and social factors, including area-level indicators of socioeconomic status, to identify factors associated with meningococcal carriage.RESULTS:In a multivariable log binomial regression model adjusted for age, lower socioeconomic status was significantly associated with higher prevalence of meningococcal carriage. A 1-unit increase in Index of Multiple Deprivation (2010) score was associated with a 1.7% increase in meningococcal carriage prevalence (95% confidence interval 0.3-3.0%). Age was the only significant predictor of carriage of Neisseria lactamica.CONCLUSIONS:Living in a deprived area is associated with increased carriage of Group B meningococcus. Deprivation is an important factor to consider in the evaluation of the effectiveness and cost-effectiveness of the introduction of new meningococcal B vaccines and the development and implementation of immunization policies. Further work is required to understand whether deprivation has an effect on meningococcal carriage through other factors such as smoking.
Using a short pulse width x-ray source and measuring the time-of-flight of photons that scatter from an object under inspection allows for the point of interaction to be determined, and a profile of the object to be sampled along the path of the beam. A three dimensional image can be formed by interrogating the entire object. Using high energy x rays enables the inspection of cargo containers with steel walls, in the search for concealed items. A longer pulse width x-ray source can also be used with deconvolution techniques to determine the points of interaction. We present time-of-flight results from both short (picosecond) width and long (hundreds of nanoseconds) width x-ray sources, and show that the position of scatter can be localised with a resolution of 2 ns, equivalent to 30 cm, for a 3 cm thick plastic test object.