Objective. The radiation response of alanine is very well characterized in the MV photon energy range where it can be used to determine the dose delivered with an accuracy better than 1%, making it suitable as a secondary standard detector in cancer radiation therapy. This is not the case in the very low energy keV x-ray range where the alanine response is affected by large uncertainties and is strongly dependent on the x-ray beam energy. This motivated the study undertaken here.Approach. Alanine pellets with a nominal thickness of 0.5 mm and diameter of 5 mm were irradiated with monoenergetic x-rays at the Diamond Light Source synchrotron, to quantify their response in the 8-20 keV range relative to60Co radiation. The absorbed dose to graphite was measured with a small portable graphite calorimeter, and the DOSRZnrc code in the EGSnrc Monte Carlo package was used to calculate conversion factors between the measured dose to graphite and the absorbed dose to water delivered to the alanine pellets. GafChromic EBT3 films were used to measure the beam profile for modelling in the MC simulations.Main results. The relative responses measured in this energy range were found to range from 0.616 to 0.643, with a combined relative expanded uncertainty of 3.4%-3.5% (k= 2), where the majority of the uncertainty originated from the uncertainty in the alanine readout, due to the small size of the pellets used.Significance. The measured values were in good agreement with previously published data in the overlapping region of x-ray energies, while this work extended the dataset to lower energies. By measuring the response to monoenergetic x-rays, the response to a more complex broad-spectrum x-ray source can be inferred if the spectrum is known, meaning that this work supports the establishment of alanine as a secondary standard dosimeter for low-energy x-ray sources.
Objective. To report the use of a portable primary standard level graphite calorimeter for direct dose determination in clinical pencil beam scanning proton beams, which forms part of the recommendations of the proposed Institute of Physics and Engineering in Medicine (IPEM) Code of Practice (CoP) for proton therapy dosimetry.Approach. The primary standard proton calorimeter (PSPC) was developed at the National Physical Laboratory (NPL) and measurements were performed at four clinical proton therapy facilities that use pencil beam scanning for beam delivery. Correction factors for the presence of impurities and vacuum gaps were calculated and applied, as well as dose conversion factors to obtain dose to water. Measurements were performed in the middle of 10 × 10 × 10 cm3homogeneous dose volumes, centred at 10.0, 15.0 and 25.0 g·cm-2depth in water. The absorbed dose to water determined with the calorimeter was compared to the dose obtained using PTW Roos-type ionisation chambers calibrated in terms of absorbed dose to water in60Co applying the recommendations in the IAEA TRS-398 CoP.Main results.The relative dose difference between the two protocols varied between 0.4% and 2.1% depending on the facility. The reported overall uncertainty in the determination of absorbed dose to water using the calorimeter is 0.9% (k= 1), which corresponds to a significant reduction of uncertainty in comparison with the TRS-398 CoP (currently with an uncertainty equal or larger than 2.0% (k= 1) for proton beams).Significance. The establishment of a purpose-built primary standard and associated CoP will considerably reduce the uncertainty of the absorbed dose to water determination and ensure improved accuracy and consistency in the dose delivered to patients treated with proton therapy and bring proton reference dosimetry uncertainty in line with megavoltage photon radiotherapy.
Background and purpose: Accurate dosimetry in Ultra-High Dose Rate (UHDR) beams is challenging because high levels of ion recombination occur within ionisation chambers used as reference dosimeters. A Small-body Portable Graphite Calorimeter (SPGC) exhibiting a dose-rate independent response was built to offer reduced uncertainty on secondary standard dosimetry in UHDR regimes. The aim of this study was to quantify the effect of the geometry and material properties of the device on the dose measurement. Materials and methods: A detailed model of the SPGC was built in the Monte Carlo code TOPAS (v3.6.1) to derive the impurity and gap correction factors, kimp and kgap. A dose conversion factor, DMC using FLUKA (v2021.2.0). These factors convert the average dose to its graphite core to the dose-to-water for a 249.7 MeV mono-energetic spot-scanned clinical proton beam. The effect of the surrounding Styrofoam on the dose measurement was examined in the simulations by substituting it for graphite. Results: The kimp and kgap correction factors were 0.9993 +/- 0.0002 and 1.0000 +/- 0.0001, respectively when the Styrofoam was not substituted, and 1.0037 +/- 0.0002 and 0.9999 +/- 0.0001, respectively when substituted for graphite. The dose conversion factor was calculated to be 1.0806 +/- 0.0001. All uncertainties are Type A. Conclusions: Impurity and gap correction factors, and the dose conversion factor were calculated for the SPGC in a FLASH proton beam. Separating out the effect of scatter from Styrofoam insulation showed this as the dominating correction factor, amounting to 1.0043 +/- 0.0002.
Objective. A calibration service based on a primary standard calorimeter for the direct determination of absorbed dose for proton beams does not exist. A new Code of Practice (CoP) for reference dosimetry of proton beams is being developed by a working party of the UK Institute of Physics and Engineering in Medicine (IPEM), which will recommend that ionisation chambers are calibrated directly in their clinical beams against the proposed Primary Standard Proton Calorimeter (PSPC) developed at the National Physical Laboratory (NPL). The aim of this work is to report on the use of the NPL PSPC to directly calibrate ionisation chambers in a low-energy passively scattered proton beam following recommendations of the upcoming IPEM CoP.Approach. A comparison between the dose derived using the proposed IPEM CoP and the IAEA TRS-398 protocol was performed, andkQvalues were determined experimentally for three types of chambers. In total, 9 plane-parallel and 3 cylindrical chambers were calibrated using the two protocols for two separate visits.Main results. The ratio of absorbed dose to water obtained with the PSPC and with ionisation chambers applying TRS-398 varied between 0.98 and 1.00, depending on the chamber type. The new procedure based on the PSPC provides a significant improvement in uncertainty where absorbed dose to water measured with a user chamber is reported with an uncertainty of 0.9% (1σ), whereas the TRS-398 protocol reports an uncertainty of 2.0% and 2.3% (1σ) for cylindrical and plane-parallel chambers, respectively. ThekQvalues found agree within uncertainties with those from TRS-398 and Monte Carlo calculations.Significance. The establishment of a primary standard calorimeter for the determination of absorbed dose in proton beams combined with the introduction of the associated calibration service following the IPEM recommendations will reduce the uncertainty and improve consistency in the dose delivered to patients.
The objective of this work was to measure the ion recombination correction for parallel-plate ionization chambers in carbon ion beams. As confirmed by the literature, due to the high LET of the beams, we have shown that the ion recombination mechanism is dominated by initial recombination. Significant differences appear depending on the orientation of the chamber.
Introduction: Reference dosimetry of broad passively scattered light-ion beams using calorimeters and ionization chambers has been widely reported and well-established. For scanned beams, however, only very few reports on the application of calorimeters exist and current dosimetry protocols do not address some important aspects of ionization chamber based dosimetry, such as the need for reference dosimetry at the entrance, the effect of complex time dependent dose deposition patterns on calorimeter and ionization chamber response and the effect of high-instantaneous dose rates on charge recombination in ionization chambers. There is also little evidence that detectors used for relative dosimetry behave similarly in scanned beams as in passively scattered beams.
Developments in hadron therapy require efforts to improve the accuracy of the dose delivered to a target volume. Here, the determination of the absorbed dose under reference conditions was analysed. Based on the International Atomic Energy Agency TRS-398 code of practice, for hadron beams, the combined standard uncertainty on absorbed dose to water under reference conditions, derived from ionisation chambers, is too large. This uncertainty is dominated by the beam quality correction factors, [Formula: see text], mainly due to the mean energy to produce one ion pair in air, wair. A method to reduce this uncertainty is to carry out primary dosimetry, using calorimetry. A [Formula: see text]-value can be derived from a direct comparison between calorimetry and ionometry. Here, this comparison is performed using a graphite calorimeter in an 80-MeV A(-1) carbon ion beam. Assuming recommended TRS-398 values of water-to-graphite stopping power ratio and the perturbation factor for an ionisation chamber, preliminary results indicate a wair-value of 35.5 ± 0.9 J C(-1).
Purpose: To reduce the uncertainty of the beam quality correction factor kQ,Q0, for scattered proton beams (SPB). This factor is used in dosimetry protocols, to determine absorbed dose-to-water with ionization chambers. For the Roos plane parallel chambers (RPPICs), the IAEA TRS-398 protocol estimates kQ,Q0-factor to be 1.004(for a beam quality Rres=2 g.cm2), with an uncertainty of 2.1%. Methods: A graphite calorimeter (GCal), a water calorimeter (WCal) and RPPICs were exposed, in a single experiment, to a 60 MeV non-modulated SPB. RPPICs were calibrated in terms of absorbed dose-to-water in a 20 MeV electron beam. The calibration coefficient is traceable to NPL's absorbed dose standards. Chamber measurements were corrected for environmental conditions, recombination and polarity. The WCal corrections include heat loss, heat defect and vessel perturbation. The GCal corrections include heat loss and absorbed dose conversion. Except for heat loss correction and its uncertainty in the WCal system, all major corrections were included in the analysis. Other minor corrections, such as beam profile non-uniformity, are still to be evaluated. Experimental kQ,Q0-factors were derived by comparing the results obtained with both calorimeters and ionometry. Results: The absorbed dose-to-water from both calorimeters was found to be within 1.3% with an uncertainty of 1.2%. kQ,Q0-factor for a RPPIC was found to be 0.998 and 1.011, with a standard uncertainty of 1.4% and 0.9% when the dose is based on the GCal and the WCal, respectively. Conclusion: Results suggest the possibility to determine kQ,Q0-values for PPICs in SPB with a lower uncertainty than specified in the TRS-398 thereby helping to reduce uncertainty on absorbed dose-to-water. The agreement between calorimeters confirms the possibility to use GCal or WCal as primary standard in SPB. Because of the dose conversion, the use of GCal may lead to slightly higher uncertainty, but is, at present, considerably easier to operate.
Purpose: Comparison between the response of a primary standard graphite calorimeter and the response of an ionization chamber (IC) allows the estimations of values of the beam quality correction factor, kQ,Q0, for the IC. This factor is used in the IAEA TRS‐398 dosimetric protocol to derive the absorbed dose‐to‐water using an IC. It depends on three parameters, one of which is the average energy required to produce an ion pair in dry air (Wair). The aim of this work is the experimental determination of the Wair‐value for an 80 MeV/A carbon ion beam. Methods: The dose measured with a graphite calorimeter, after correction for heat transfer between different parts of the calorimeter, must be converted to dose‐to‐water. This conversion needs knowledge of the water‐to‐graphite stopping power ratio and the fluence correction factor. These have been determined experimentally and numerically using Geant4 Monte Carlo simulations. In addition to these calorimetric corrections, the determination of kQ,Q0‐values needs the IC reading to be corrected for ion recombination. Results: The water‐to‐graphite stopping power ratio has been determined numerically to be 1.115 while the fluence correction factor has been determined experimentally and numerically to vary linearly between 0.9996 at the phantom surface and 1.0063 before the Bragg peak. The analysis of the ion recombination correction measurements is still ongoing. Two experimental campaigns with the graphite calorimeter and several IC's have been performed. Based on recommended data from TRS‐398, we derived preliminary Wair‐values of 36. J/C and 34.85 J/C for each campaign. These results are under investigation. Conclusion: These results demonstrate the feasibility of graphite calorimetry in an 80 MeV/A carbon ion beam and are encouraging for our future work: a direct comparison between water and graphite calorimetry in a clinical carbon ion beam. Funding support for Severine Rossomme: BioWin program of the Walloon Government (Belgium); Funding support for NPL people: Acoustics and Ionising Radiation Metrology Programme of the National Measurement System (UK)
Based on experiments and numerical simulations, a study is carried out pertaining to the conversion of dose-to-graphite to dose-to-water in a carbon ion beam. This conversion is needed to establish graphite calorimeters as primary standards of absorbed dose in these beams. It is governed by the water-to-graphite mass collision stopping power ratio and fluence correction factors, which depend on the particle fluence distributions in each of the two media. The paper focuses on the experimental and numerical determination of this fluence correction factor for an 80 MeV/A carbon ion beam. Measurements have been performed in the nuclear physics laboratory INFN-LNS in Catania (Sicily, Italy). The numerical simulations have been made with a Geant4 Monte Carlo code through the GATE simulation platform. The experimental data are in good agreement with the simulated results for the fluence correction factors and are found to be close to unity. The experimental values increase with depth reaching 1.010 before the Bragg peak region. They have been determined with an uncertainty of 0.25%. Different numerical results are obtained depending on the level of approximation made in calculating the fluence correction factors. When considering carbon ions only, the difference between measured and calculated values is maximal just before the Bragg peak, but its value is less than 1.005. The numerical value is close to unity at the surface and increases to 1.005 near the Bragg peak. When the fluence of all charged particles is considered, the fluence correction factors are lower than unity at the surface and increase with depth up to 1.025 before the Bragg peak. Besides carbon ions, secondary particles created due to nuclear interactions have to be included in the analysis: boron ions ((10)B and (11)B), beryllium ions ((7)Be), alpha particles and protons. At the conclusion of this work, we have the conversion of dose-to-graphite to dose-to-water to apply to the response of a graphite calorimeter in an 80 MeV/A carbon ion beam. This conversion consists of the product of two contributions: the water-to-graphite electronic mass collision stopping power ratio, which is equal to 1.115, and the fluence correction factor which varies linearly with depth, as k(fl, all) = 0.9995 + 0.0048(zw-eq). The latter has been determined on the basis of experiments and numerical simulations.
PURPOSE:The IAEA TRS-398 code of practice can be applied for the measurement of absorbed dose to water under reference conditions with an ionization chamber. For protons, the combined relative standard uncertainty on those measurements is less than 2% while for light-ion beams, it is considerably larger, i.e. 3.2%, mainly due to the higher uncertainty contributions for the water to air stopping power ration and the W air-value on the beam quality correction factors kQ,Q0 . To decrease this uncertainty, a quantification of kQ,Q0 is proposed using a primary standard level graphite calorimeter. This work includes numerical and experimental determinations of dose conversion factors to derive dose to water from graphite calorimetry. It also reports on the first experimental data obtained with the graphite calorimeter in proton, alpha and carbon ion beams.METHODS:Firstly, the dose conversion has been calculated with by Geant4 Monte-Carlo simulations through the determination of the water to graphite stopping power ratio and the fluence correction factor. The latter factor was also derived by comparison of measured ionization curves in graphite and water. Secondly, kQ,Q0 was obtained by comparison of the dose response of ionization chambers with that of the calorimeter.RESULTS:Stopping power ratios are found to vary by no more than 0.35% up to the Bragg peak, while fluence correction factors are shown to increase slightly above unity close to the Bragg peak. The comparison of the calorimeter with ionization chambers is currently under analysis. For the modulated proton beam, preliminary results on W air confirm the value recommended in TRS-398. Data in both the non-modulated proton and light-ion beams indicate higher values but further investigation of heat loss corrections is needed.CONCLUSIONS:The application of graphite calorimetry to proton, alpha and carbon ion beams has been demonstrated successfully. Other experimental campaigns will be held in 2012. This work is supported by the BioWin program of the Wallon Government.