Detecting vapor from drugs and explosives is challenging due to their extremely low vapor pressures, often in the parts-per-trillion (pptv) to parts-per-quadrillion (ppqv) range. Atmospheric flow tube-mass spectrometry (AFT-MS) overcomes this limitation by extending ion-molecule reaction times, achieving detection of ppqv levels of vapor and femtogram levels of residue. However, the prior AFT-MS system was large (∼435 lbs.), limiting field deployment. To enhance portability, the AFT was integrated with a miniaturized, ruggedized, field-portable linear ion trap mass spectrometer (∼45 lbs.). This work presents results demonstrating the feasibility of portable AFT-MS for detecting illicit drugs and explosives in the field. Initial tests of the AFT with the portable MS were conducted to detect trace vapors from both solutions and residues. Real-time detection of fentanyl vapor from residue indicates detection levels of less than 100 ppqv based on the equilibrium vapor pressure of fentanyl. Adjusting ionization parameters, sampling speeds, trapping times, pressure levels, and system timing revealed pathways for further sensitivity enhancements. Early field tests at a U.S. Customs and Border Protection port of entry confirmed the ability of the system to simultaneously detect multiple drugs in complex mixtures. This is the first demonstration of portable vapor detection with AFT-MS, achieving subpptv-level sensitivity for real-time field detection of drugs and explosives.
Abstract Mass spectrometry is used to identify chemicals to which humans are exposed, but it cannot directly determine molecular structures. Instead, structures are inferred by matching experimental spectra to libraries of spectra constructed from analyses of pure reference compounds. However, the chemical space of human exposures far exceeds the amount of experimental library spectra. Here, we evaluate a ‘reference-free’ strategy for confident identification of unknown molecules. Using fentanyl as a case study, we created a suspect library of over 1 billion computationally predicted fentanyl analogs and predicted molecular properties through machine learning, molecular dynamics, and density functional theory. Multi-dimensional spectra from a blinded analysis of a mock fentanyl tablet were matched with the predicted library, yielding an average of three candidate structures per measured analog, with six exact identifications. This work emphasizes the promise of reference-free molecular measurements for assessing human exposure by merging computational predictions with high-dimensional measurements.
This study demonstrates ion focusing at ambient pressure and increased ion signal by creating a voltage difference from a field-free region to a detector, thereby improving the detection of chemicals such as explosives and drugs. At ambient pressure, ion loss and the resulting signal reduction pose challenges that limit detection sensitivity in analytical instruments. Techniques to increase sensitivity, such as atmospheric flow tube-mass spectrometry (AFT-MS), extend ion-molecule reaction times but result in significant overall ion loss due to diffusion. Ion manipulation techniques, though challenging at ambient pressure, can mitigate these losses by concentrating ions toward the detector inlet. Using SIMION, electric fields and ion trajectories were modeled with a voltage difference applied between a flow tube and a detector, revealing the potential for ion focusing at ambient pressure. Experimentation with an atmospheric flow tube, a mass spectrometer, and Faraday plate detectors was used to measure ion beam profiles across varying flow rates, tube diameters, and voltage differences. Application of a voltage difference effectively directed ions to the axial center of the flow tube, narrowed ion beam width, and increased signal intensity by 5 to 10 times for mass spectrometer and Faraday plate detectors, respectively, compared to conditions without a voltage difference. This ion-focusing approach shows promise for improving sensitivity in ambient-pressure instruments. This technique has the potential to enhance detection levels in security and forensic applications, with particular benefits for field-portable devices that can be used at checkpoints to identify explosives and drugs.
Identification of compounds with minimal ambiguity remains a central challenge in mass spectrometry-based metabolomics. Conventional compound identification relies on comparing analytical signatures (e.g., mass-to-charge ratio, collision cross section, tandem mass spectra) against reference data obtained from measurements of authentic chemical standards. The breadth of annotatable compounds using this approach is necessarily limited by availability of authentic standards, analytical throughput, and resolving power of the separations that underly the measurements. The maturation of computational methods, both theory-driven and artificial intelligence/machine learning-based, for prediction of various molecular properties relevant to multidimensional mass spectrometry measurements has opened the door to a new "reference-free" paradigm of compound annotation. Through augmenting existing reference data for molecular properties with computational predictions, the universe of identifiable chemical species can be expanded significantly beyond its current limits. An unexplored aspect of this novel approach is understanding how to gauge confidence in resulting annotations, especially as the compound search space is expanded. Intuitively, the confidence of a compound annotation is related to the inherent discriminatory power of the molecular properties used for identification, as well as the precision with which the properties are measured or predicted. In this work, we characterize this relationship between measurement precision and identification probability in a systematic and quantitative fashion for a defined region of chemical space that includes organic small molecule metabolites. Importantly, this work establishes a framework for conducting metabolite identification probability analysis that enables others to quantify this relationship for their own compounds and properties of interest.
The vapor pressures of many explosive compounds and illicit drugs are extremely low relative to volatile organic compounds. Determining accurate vapor pressure values at ambient temperature can be difficult. Vapor pressure measurements at higher temperatures are often used with vapor pressure equations to extrapolate the vapor pressure to 25 °C. Many researchers have measured and reported vapor pressure values for explosives or drugs, and there can be large variations (orders of magnitude) in reported vapor pressure values. Advances in sensitivity of vapor detection technologies has enabled measurement of the vapor pressure of low-volatility compounds at lower temperatures resulting in updated vapor pressure values in the literature. Knowing accurate vapor pressures of compounds is important to understand the feasibility of detecting substances for a variety of security applications.This review of experimentally-measured vapor pressures of explosives and drugs is intended to be a companion to the prior review in TrAC by Ewing et al. (2013). Updates are provided on reported vapor pressure values of various explosives since the prior review. The content is expanded to include some vapor pressure values of drugs with the same intent as to provide benchmark levels in estimating vapor concentrations that could be available for detection. Vapor pressure relationships traceable to the original citations, including the temperature ranges for the measurements, are compiled. After critical examination of the available data, the vapor pressure values at 25 °C are tabulated and summarized in a figure, which displays the relationship of vapor pressures between various explosives and drugs.
2-Benzylbenzimidazoles, or "nitazenes", are a class of novel synthetic opioids (NSOs) that are increasingly being detected alongside fentanyl analogs and other opioids in drug overdose cases. Nitazenes can be 20× more potent than fentanyl but are not routinely tested for during postmortem or clinical toxicology drug screens; thus, their prevalence in drug overdose cases may be under-reported. Traditional analytical workflows utilizing liquid chromatography-tandem mass spectrometry (LC-MS/MS) often require additional confirmation with authentic reference standards to identify a novel nitazene. However, additional analytical measurements with ion mobility spectrometry (IMS) may provide a path toward reference-free identification, which would greatly accelerate NSO identification rates in toxicology laboratories. Presented here are the first IMS and collision cross section (CCS) measurements on a set of fourteen nitazene analogs using a structures for lossless ion manipulations (SLIM)-orbitrap MS. All nitazenes exhibited two high intensity baseline-separated IMS distributions, which fentanyls and other drug and druglike compounds also exhibit. Incorporating water into the electrospray ionization (ESI) solution caused the intensities of the higher mobility IMS distributions to increase and the intensities of the lower mobility IMS distributions to decrease. Nitazenes lacking a nitro group at the R1 position exhibited the greatest shifts in signal intensities due to water. Furthermore, IMS-MS/MS experiments showed that the higher mobility IMS distributions of all nitazenes possessing a triethylamine group produced fragment ions with m/z 72, 100, and other low intensity fragments while the lower mobility IMS distributions only produced fragment ions with m/z 72 and 100. The IMS, solvent, and fragmentation studies provide experimental evidence that nitazenes potentially exhibit three gas-phase protomers. The cyclic IMS capability of SLIM was also employed to partially resolve four sets of structurally similar nitazene isomers (e.g., protonitazene/isotonitazene, butonitazene/isobutonitazene/secbutonitazene), showcasing the potential of using high-resolution IMS separations in MS-based workflows for reference-free identification of emerging nitazenes and other NSOs.
Vapor detection is a noncontact sampling method, which is a less invasive means of explosives screening than physical swiping. Explosive vapor detection is a challenge due to the low levels of vapors available for detection. This study demonstrates that the parts-per-quadrillion sensitivity of atmospheric flow tube-mass spectrometry (AFT-MS) combined with a high-volume air sampler enables standoff detection of trace explosives vapor at distances of centimeters to meters. Standoff detection of explosives vapor was possible both upstream and downstream of the vapor source relative to room air currents. RDX vapor from a saturated source was detected at up to 2.5 m. Vapors from RDX residue and nitroglycerin residue were detected at distances up to 0.5 m. The sampling can be optimized by accounting for air movement in the room or environment, which could further extend standoff detection distances. Using AFT-MS with a high-volume sampler could also be effective for standoff vapor detection of drugs and additional chemical threats and could be useful for security screening applications such as at mail facilities, border crossings, and security checkpoints.
The opioid crisis in the United States is being fueled by the rapid emergence of new fentanyl analogs and precursors that can elude traditional library-based screening methods, which require data from known reference compounds. Since reference compounds are unavailable for new fentanyl analogs, we examined if fentanyls (fentanyl + fentanyl analogs) could be identified in a reference-free manner using a combination of electrospray ionization (ESI), high-resolution ion mobility (IM) spectrometry, high-resolution mass spectrometry (MS), and higher-energy collision-induced dissociation (MS/MS). We analyzed a mixture containing nine fentanyls and W-15 (a structurally similar molecule) and found that the protonated forms of all fentanyls exhibited two baseline-separated IM distributions that produced different MS/MS patterns. Upon fragmentation, both IM distributions of all fentanyls produced two high intensity fragments, resulting from amine site cleavages. The higher mobility distributions of all fentanyls also produced several low intensity fragments, but surprisingly, these same fragments exhibited much greater intensities in the lower mobility distributions. This observation demonstrates that many fragments of fentanyls predominantly originate from one of two different gas-phase structures (suggestive of protomers). Furthermore, increasing the water concentration in the ESI solution increased the intensity of the lower mobility distribution relative to the higher mobility distribution, which further supports that fentanyls exist as two gas-phase protomers. Our observations on the IM and MS/MS properties of fentanyls can be exploited to positively differentiate fentanyls from other compounds without requiring reference libraries and will hopefully assist first responders and law enforcement in combating new and emerging fentanyls.
On August 9-10, 2023, a workshop was convened at the Pacific Northwest National Laboratory (PNNL) in Richland, WA that brought together a group of internationally recognized experts in metabolomics, natural products discovery, chemical ecology, chemical and biological threat assessment, cheminformatics, computational chemistry, cloud computing, artificial intelligence, and novel technology development. These experts were invited to assess the value and feasibility of a grand-scale project to create new technologies that would allow the identification and quantification of all small molecules, or to decode the molecular universe. The Decoding the Molecular Universe project would extend and complement the success of the Human Genome Project by developing new capabilities and technologies to measure small molecules (defined as non-protein, non-polymer molecules less than 1500 Daltons) of any origin and generated in biological systems or produced abiotically. Workshop attendees 1) explored what new understanding of biological and environmental systems could be revealed through the lens of small molecules; 2) characterized the similarities in current needs and technical challenges between each science or mission area for unambiguous and comprehensive determination of the composition and quantities of small molecules of any sample; 3) determined the extent to which technologies or methods currently exist for unambiguously and comprehensively determining the small molecule composition of any sample and in a reasonable time; and 4) identified the attributes of the ideal technology or approach for universal small molecule measurement and identification. The workshop concluded with a discussion of how a project of this scale could be undertaken, possible thrusts for the project, early proof-of-principle applications, and similar efforts upon which the project could be modeled.
High-resolution ion mobility spectrometry-mass spectrometry (HR-IMS-MS) instruments have enormously advanced the ability to characterize complex biological mixtures. Unfortunately, HR-IMS and HR-MS measurements are typically performed independently due to mismatches in analysis time scales. Here, we overcome this limitation by using a dual-gated ion injection approach to couple an 11 m path length structures for lossless ion manipulations (SLIM) module to a Q-Exactive Plus Orbitrap MS platform. The dual-gate setup was implemented by placing one ion gate before the SLIM module and a second ion gate after the module. The dual-gated ion injection approach allowed the new SLIM-Orbitrap platform to simultaneously perform an 11 m SLIM separation, Orbitrap mass analysis using the highest selectable mass resolution setting (up to 140 k), and high-energy collision-induced dissociation (HCD) in ∼25 min over an m/z range of ∼1500 amu. The SLIM-Orbitrap platform was initially characterized using a mixture of standard phosphazene cations and demonstrated an average SLIM CCS resolving power (RpCCS) of ∼218 and an SLIM peak capacity of ∼156, while simultaneously obtaining high mass resolutions. SLIM-Orbitrap analysis with fragmentation was then performed on mixtures of standard peptides and two reverse peptides (SDGRG1+, GRGDS1+, and RpCCS = 305) to demonstrate the utility of combined HR-IMS-MS/MS measurements for peptide identification. Our new HR-IMS-MS/MS capability was further demonstrated by analyzing a complex lipid mixture and showcasing SLIM separations on isobaric lipids. This new SLIM-Orbitrap platform demonstrates a critical new capability for proteomics and lipidomics applications, and the high-resolution multimodal data obtained using this system establish the foundation for reference-free identification of unknown ion structures.
On August 9-10, 2023, a workshop was convened at the Pacific Northwest National Laboratory (PNNL) in Richland, WA that brought together a group of internationally recognized experts in metabolomics, natural products discovery, chemical ecology, chemical and biological threat assessment, cheminformatics, computational chemistry, cloud computing, artificial intelligence, and novel technology development. These experts were invited to assess the value and feasibility of a grand-scale project to create new technologies that would allow the identification and quantification of all small molecules, or to decode the molecular universe. The Decoding the Molecular Universe project would extend and complement the success of the Human Genome Project by developing new capabilities and technologies to measure small molecules (defined as non-protein, non-polymer molecules less than 1500 Daltons) of any origin and generated in biological systems or produced abiotically. Workshop attendees 1) explored what new understanding of biological and environmental systems could be revealed through the lens of small molecules; 2) characterized the similarities in current needs and technical challenges between each science or mission area for unambiguous and comprehensive determination of the composition and quantities of small molecules of any sample; 3) determined the extent to which technologies or methods currently exist for unambiguously and comprehensively determining the small molecule composition of any sample and in a reasonable time; and 4) identified the attributes of the ideal technology or approach for universal small molecule measurement and identification. The workshop concluded with a discussion of how a project of this scale could be undertaken, possible thrusts for the project, early proof-of-principle applications, and similar efforts upon which the project could be modeled.
Increasing ion molecule reaction times for ambient ionization techniques can increase sensitivity of detection. Longer reaction times result in an increase in analyte signal relative to the reactant ion signal, however with a subsequent decrease in the total ion intensity. This loss in the total number of ions reaching the detector limits the extent to which increased reaction time can improve detection in practical applications. In this study ion loss was measured using either electric fields or gas flow to control ion transit time. Ion transit times ranged from 40 ms to 8 s, which resulted in ion densities ranging from 5 x 106 to 2 x 104 ions/cm3, respectively. These results provide insights into practical reaction time limitations or experimental boundaries when exploring sensitivity enhancements in ambient ionization. When exploring longer reaction times with an atmospheric flow tube coupled to a mass spectrometer (AFT-MS), an additional corona discharge ionization source of the same polarity was added in an attempt to increase total ion signal, but no improvement in signal was observed. This obser-vation suggested that there was a maximum ion density being reached by the corona discharge that could not be increased by adding more of the same polarity ions. A hypothesis was that intermingling both polarities of ions would reduce ion loss and increase the measured ion signal. When two corona discharge sources-one each of positive and negative polarity-were used simultaneously in an AFT-MS, the total ion signal of either polarity measured by the MS was approximately double the current from a single corona discharge source. An increase in the negative ion signal was also observed when-10 parts-per-trillion vapor levels of an explosive (RDX, also known as hexogen) were introduced. In this case both the nitrate reactant ions and the RDX signal more than doubled with the addition of a corona discharge source producing positive ions. When intermingling positive and negative ions, an increase in ion current was also observed when the AFT was coupled directly to a Faraday detector or to an ion mobility spectrometer (IMS). This was the first demonstration of interfacing the AFT to an IMS. The observations and developments from this work will help in creating more portable instruments with improved sensitivity for detection of chemicals.
Proton affinity is a major factor in the atmospheric pressure chemical ionization of illicit drugs. The detection of illicit drugs by mass spectrometry and ion mobility spectrometry relies on the analytes having greater proton affinities than background species. Evaluating proton affinities for fentanyl and its analogues is informative for predicting the likelihood of ionization in different environments and for optimizing the compounds' ionization and detection, such as through the addition of dopant chemicals. Herein, density functional theory was used to computationally determine the proton affinity and gas-phase basicity of 15 fentanyl compounds and several relevant molecules as a reference point. The range of proton affinities for the fentanyl compounds was from 1018 to 1078 kJ/mol. Fentanyl compounds with the higher proton affinity values appeared to form a bridge between the oxygen on the amide and the protonated nitrogen on the piperidine ring based on models and calculated bond distances. Experiments with fragmentation of proton-bound clusters using atmospheric flow tube-mass spectrometry (AFT-MS) provided estimates of relative proton affinities and showed proton affinity values of fentanyl compounds >1000 kJ/mol, which were consistent with the computational results. The high proton affinities of fentanyl compounds facilitate their detection by ambient ionization techniques in complex environments. The detection limits of the fentanyl compounds with AFT-MS are in the low femtogram range, which demonstrates the feasibility of trace vapor drug detection.
There is a need for non-contact, real-time vapor detection of drugs to combat illicit transportation and help curb the opioid epidemic. The low volatility of drugs, like fentanyl, makes room temperature vapor detection of illicit drugs challenging, but feasible by atmospheric flow tube-mass spectrometry (AFT-MS). AFT-MS is a non-contact vapor detection approach capable of ultra-trace detection of drugs, including fentanyl and its analogs at low parts-per-quadrillion (ppqv) levels. The determination of vapor pressure values of fentanyl is necessary to understand potential vapor concentrations that may be available for detection. In this paper, vapor pressures of fentanyl free base and fentanyl hydrochloride salt (a common form of the illicit drug) were measured as a function of temperature at or near ambient conditions using the transpiration (gas saturation) method and AFT-MS. Based on our measurements, the vapor pressure of fentanyl at 25 °C is 9.0 × 10-14 atm (90 ppqv), and the vapor pressure of fentanyl hydrochloride at 25 °C is 1.8 × 10-17 atm (0.018 ppqv). We also demonstrate non-contact, real-time vapor detection of fentanyl. Preconcentration of vapors can further extend the detection capabilities. The collection, desorption, and detection of fentanyl vapors at ambient conditions was demonstrated for sampling times of seconds to an hour resulting in increased signal. AFT-MS is a viable detection method of fentanyl and other drugs for screening of packages and cargo.
Many modern ion mobility (IM) and mass spectrometers (MS) operate under low pressure (<= 10 Torr) and employ high voltage radiofrequencies (RF) to provide ion confinement. Unfortunately, RF effectiveness drastically decreases as pressure increases, and few techniques for focusing ions at elevated pressures exist. Here we demonstrate a new approach for focusing ions at atmospheric pressure (AP) by applying nonlinear DC voltage sequences following quadratic and power (exponential) functions to a stacked ring ion guide. We used ion trajectory simulations to rigorously explore how ions react to nonlinear electric fields and validate the simulations with a set of ion current measurements performed at AP. Ion tra-jectory simulations show that ions initially defocus near the entrance of the device but then become intensely focused as they travel through the device. Contour plots for both nonlinear voltage sequences show electric field lines that increasingly curve inwards as a function of distance, resulting in spatial ion focusing. Experimental ion current and spot size measurements were performed at AP using a 10-cm stacked ring ion guide and a segmented Faraday cup detector. Quadratic sequences produced-5% smaller spot sizes (-22.8 mm) and-25% higher ion current compared to a linear voltage sequence (-24.0 mm). Alternatively, power sequences produced-64% smaller spot sizes (-8.7 mm), albeit with-10x lower ion current. However, both nonlinear voltage sequences produced similar ion currents at the center of the Faraday cup detector, indicating that higher ion densities are achieved when using nonlinear voltage gradients. These results demonstrate a new way to focus ions at AP, and the capabilities demonstrated here provide fundamental insights on how to keep ions inside analytical devices at elevated pressures without RF. (c) 2022 Published by Elsevier B.V.
Explosives are often used in industry, geology, mining, and other applications, but it is not always clear what remains after a detonation or the fate and transport of any residual material. The goal of this study was to determine to what extent intact molecules of high explosive (HE) compounds are detectable and quantifiable from post-detonation dust and particulates in a field experiment with varied topography. We focused on HMX (1,3,5,7-Tetranitro-1,3,5,7-tetrazocane), which is less studied in field detonation literature, as the primary explosive material and RDX (1,3,5-Trinitroperhydro-1,3,5-triazine) as the secondary material. The experiment was conducted at Site 300, Lawrence Livermore National Laboratory's Experimental Test Site, in California, USA. Two 20.4 kg and one 40.8 kg above ground explosions (primarily comprised of LX-14, an HMX-based polymer-bonded high explosive) were detonated on an open-air firing area on separate days. The complex terrain of the firing area (e.g., buildings, berm, low-height obstacles) was advantageous to study HE deposition in relation to plume dynamics. Three types of samples were collected up to 100 m away from each shot: surface swipes of aluminum plates, surface swipes of fixed objects, and filters from air samples. We used atmospheric flow tube-mass spectrometry (AFT-MS) to quantify picogram levels of molecular residue of HE material in the post-detonation dust. An aliquot of sample extract in methanol (e.g., 1 mu L of 0.5 mL) was placed onto a resistive material and then thermally desorbed into the AFT-MS. We successfully detected and quantified both HMX and RDX in many of the samples. Based on mass (pg) detected and solution dilution, we back-calculated the mass collected on the swipe or filter (ng per sample). The aerial distribution of molecular residue was consistent with the path of the plume, which was strongly determined by wind speed and direction at the time of each shot. The quantity of material detected appeared to correlate more with distance from the shot and the wind conditions than with shot size. This study demonstrates that the picogram detection levels of AFT-MS are well-suited for quantification of analytes (e.g., HMX and RDX) in environmental samples.
Knowing accurate saturated vapor pressures of explosives at ambient conditions is imperative to provide realistic boundaries on available vapor for ultra-trace detection. In quantifying vapor content emanating from low-volatility explosives, we observed discrepancies between the quantity of explosive expected based on literature vapor pressure values and the amount detected near ambient temperatures. Most vapor pressure measurements for low-volatility explosives, such as RDX (1,3,5-trinitro-1,3,5-triazinane) and HMX (1,3,5,7-tetranitro-1,3,5,7-tetrazocane), have been made at temperatures far exceeding 25 °C and linear extrapolation of these higher temperature trends appears to underestimate vapor pressures near room temperature. Our goal was to measure vapor pressures as a function of temperature closer to ambient conditions. We used saturated RDX and HMX vapor sources at controlled temperatures to produce vapors that were then collected and analyzed via atmospheric flow tube-mass spectrometry (AFT-MS). The parts-per-quadrillion (ppqv) sensitivity of AFT-MS enabled measurement of RDX vapor pressures at temperatures as low as 7 °C and HMX vapor pressures at temperatures as low as 40 °C for the first time. Furthermore, these vapor pressures were corroborated with analysis of vapor generated by nebulizing low concentration solutions of RDX and HMX. We report updated vapor pressure values for both RDX and HMX. Based on our measurements, the vapor pressure of RDX at 25 °C is 3 ± 1 × 10-11 atm (i.e., 30 parts per trillion by volume, pptv), the vapor pressure of HMX is 1.0 ± 0.6 × 10-14 atm (10 ppqv) at 40 °C and, with extrapolation, HMX has a vapor pressure of 1.0 ± 0.6 × 10-15 atm (1.0 ppqv) at 25 °C.
Atmospheric flow tube-mass spectrometry (AFT-MS) enables real-time, non-contact detection of fentanyl, heroin, and other illicit drugs at parts-per-quadrillion vapor levels or from femtograms of residue.
Using a commercial mass spectrometer interfaced with an atmospheric flow tube (AFT) allowed for the detection of a variety of inorganic compounds used as oxidizers in homemade explosives (HMEs) at picogram levels. The AFT provides reaction times of between 3 and 5 s with flows of 6 L/min, enabling detection levels, after thermal desorption, similar to those previously demonstrated for RDX vapor in the low parts per quadrillion range. The thermal desorption of chlorate and perchlorate salts resulted in the production of the corresponding anions which have higher electron affinities than that of the nitrate reactant ions. A dielectric barrier discharge, used as the ionization source, produced the nitrate reactant ions. In some instances, the molecular salt formed adducts with the nitrate, chlorate, and/or perchlorate anions, giving insight into the original identity of the salt cation. Urea nitrate, guanidine nitrate, and potassium nitrate were also detected as adducts with the nitrate reactant ion. The direct room-temperature vapor detection of urea nitrate and hydrogen peroxide, which have relatively high vapor pressures compared to the other salts in this study, is also demonstrated. Room-temperature vapor detection of chlorate and perchlorate salts is possible by the addition of a dilute acid which converts the salt into a more volatile acidic form. A discussion of the instrumentation, methods used, and the ionization chemistry is provided.