Cemented unicompartmental knee arthroplasty (UKA) shows good survivorship and function. However, implant failure, causing the need for revision, can occur. Aseptic loosening is still among the most common reasons for revision. The purpose of this study was to assess the influence of preimplantation lavage technique on tibial cement penetration depth, tibial cement volume, and load to fracture in the tibial component of mobile-bearing UKA. In 10 pairs of fresh frozen human tibiae, cemented UKA was implanted by an experienced surgeon. Tibial components were then implanted, left and right tibiae were randomly allocated to group A or B. Prior to implantation, irrigation was performed with either syringe lavage or pulsatile jet lavage in a standardized manner. Cement surface was 4170.2mm (2) (3271.6-5497.8mm (2) ) in the syringe lavage group, whereas the jet lavage group showed 4499.3mm (2) (3354.3-5809.1mm (2) ); cement volume was significantly higher as well (4143.4mm (3) (2956.6-6198.6mm (3) ) compared with 5936.9mm (3) (3077.5-8183.1mm (3) )). Cement penetration depth was 2.5mm (1.7-3.2mm) for the jet lavage, and 1.8mm (1.2-2.4mm) for the syringe lavage. The mean fracture load was 4680N in the jet lavage group and 3800N in the syringe lavage group ( p =0.001). Subsidence was significantly higher for syringe lavage. This study suggests a correlation of cement penetration depth and cement volume to implant failure in the tibial component of a UKA using a cadaveric model. The type of bone lavage most likely influences these two key parameters.
Aims The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation. Methods Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system. Results At the main fixation zone, the twin peg shows less relative movement at 70°/115°. At the transition zone, relative movements are smaller for the single peg for both angles. The single peg shows higher compression at 70° flexion, whereas the twin peg design shows higher compression at 115°. X-displacement is significantly higher for the single peg at 115°. Conclusion Bony defects should be avoided in OUKA. The twin peg shows high resilience against push-out force and should be preferred over the single peg. Cite this article: Bone Joint Res 2022;11(2):82–90.
In cemented joint arthroplasty, state-of-the-art cementing techniques include high-pressure pulsatile saline lavage prior to cementation. Even with its outstanding importance in cementation, there are surprisingly few studies regarding the physical parameters that define pulsatile lavage systems. To investigate the parameters of impact pressure, flow rate, frequency and the cleaning effect in cancellous bone, we established a standardized laboratory model. Standardized fat-filled carbon foam specimens representing human cancellous bone were cleaned with three different high-pressure pulsatile lavage systems. Via CT scans before and after cleaning, the cleaning effect was evaluated. All systems showed a cleaning depth of at least 3.0 mm and therefore can be generally recommended to clean cancellous bone in cemented joint arthroplasty. When comparing the three lavage systems, the study showed significant differences regarding cleaning depths and volume, with one system being superior to its peer systems. Regarding the physical parameters, high impact pressure in combination with high flow rate and longer distance to the flushed object seems to be the best combination to improve the cleaning of cancellous bone and therefore increase the chances of a deeper cement penetration that is required in cemented joint arthroplasty. In summary, this study provides the first standardized comparison of different lavage systems and thus gives initial guidance on how to optimally prepare cancellous bone for cemented joint arthroplasty.
Neuropeptide Y (NPY), a sympathetic cotransmitter, has both prejunctional and postjunctional actions in the cardiovascular system. In the bioassay system used here NPY attenuates cardiac vagal action (an indicator of prejunctional or Y2 action) and increases blood pressure (an indicator of postjunctional or Y1 action). [Leu31,Pro34]NPY has little or no prejunctional activity i.e does not attenuate cardiac vagal action, but increases blood pressure as effectively as NPY. [Leu31,Pro34]NPY can therefore be used in functional experiments to distinguish between the two types of NPY receptor.
Aim:To compare changes in apparent diffusion coefficient (ADC) in neonatal meningitis using serial diffusion-weighted imaging (DWI).Method:Thirty neonates with meningitis and 12 age/sex-matched controls were studied using DWI. ADC was quantified by placing region of interest(s) on periventricular white matter during acute illness and again at 21 days. Three groups of patients were studied: those with normal findings on both conventional MRI and DWI, those with abnormal DWI only and those with abnormal conventional MRI as well as DWI. Neurodevelopment assessment was performed in controls and patients at 3 months using Indian adaptation of Bayley scales of infant development (BSID) kit.Results:Patients with neonatal meningitis with normal imaging (n = 8) showed no significant difference in ADC compared to controls. Patients showing abnormality only on DWI (n = 10) and on both conventional magnetic resonance imaging (MRI) as well as DWI (n = 12) had significantly reduced ADC (p = 0.001) than controls at baseline study. Follow-up study showed no significant differences in ADC in controls compared to any patient group. Significantly reduced neurodevelopmental scores were observed in patient groups compared to controls.Conclusion:We conclude that quantitative ADC may detect meningitis-induced hypoxia early in brain parenchyma, which may be associated with abnormal motor and mental development.
Background: Aseptic loosening remains one of the most common causes of revision of the tibial component for total knee arthroplasty. A stable bond between implant and cement is essential for appropriate long-term results. The aim of our in vitro study was to investigate the maximum failure load of tibial ATTUNE prosthesis design alternatives compared with a previous design. In addition, cement-in-cement revision was considered as a potential strategy after tibial component debonding. Methods: The experimental investigations of the maximum failure load of the implant-cement interface were performed under optimal conditions, without potential contamination. We compared the designs of the tibial components of the ATTUNE, ATTUNE S+ and P.F.C. Sigma. In addition, we investigated the cement-in-cement revision for the ATTUNE knee system replacing it with an ATTUNE S+. Results: The maximum failure load showed no significant difference between P.F.C. Sigma and ATTUNE groups (P = 0.087), but there was a significant difference between the P.F.C. Sigma and the ATTUNE S+ groups (P < 0.001). The analysis also showed a significant difference (P < 0.001) between the ATTUNE and the ATTUNE S+ groups for the maximum failure load. The ATTUNE S+ cement-in-cement revision group showed a significant higher failure load (P < 0.001) compared with the P.F.C. Sigma and ATTUNE groups. No significant differences (P = 1.000) were found between the ATTUNE S+ cement-in-cement and ATTUNE S + group. Conclusion: Based on these results, we found no design-specific evidence of increased debonding risk with the ATTUNE and ATTUNE S+ components compared with the P.F.C Sigma. Furthermore, the cement-in-cement revision seems to be an alternative for the revision surgery. (c) 2021 Elsevier B.V. All rights reserved.
Acetabular cup deformation may affect liner/cup congruency, clearance and/or osseointegration. It is unclear, whether deformation of the acetabular components occurs during load and to what extent. To evaluate this, revision multi-hole cups were implanted into six cadaver hemipelvises in two scenarios: without acetabular defect (ND); with a large acetabular defect (LD) that was treated with an augment. In the LD scenario, the cup and augment were attached to the bone and each other with screws. Subsequently, the implanted hemipelvises were loaded under a physiologic partial-weight-bearing modality. The deformation of the acetabular components was determined using a best-fit algorithm. The statistical evaluation involved repeated-measures ANOVA. The mean elastic distension of the ND cup was 292.9 µm (SD 12.2 µm); in the LD scenario, 43.7 µm (SD 11.2 µm); the mean maximal augment distension was 79.6 µm (SD 21.6 µm). A significant difference between the maximal distension of the cups in both scenarios was noted (F(1, 10) = 11.404; p = 0.007). No significant difference was noted between the compression of the ND and LD cups, nor between LD cups and LD augments. The LD cup displayed significantly lower elastic distension than the ND cup, most likely due to increased stiffness from the affixed augment and screw fixation.
Background: There is little conformity regarding the surgical treatment of metastasis of the proximal femur, especially in palliative patients with limited life expectancy. Patients and Methods: En-bloc resection of secondary bone malignancies of the proximal femur and reconstruction by modular arthroplasty was performed in a consecutive series of 45 patients. The mean follow-up period was 16.4 months (0.6–74.7). Results: The survival rate of all patients was 6.6% (95% CI: 0–14.9) at 74.7 months. There was no significant difference in patients with a solitary or disseminated disease at index operation (log-rank p = 0.1214). Recurrent dislocation was the most frequent local complication (n = 6) necessitating an open reduction in four cases. The use of a Trevira tube showed a higher risk of dislocation compared to the simple bonding of remaining soft tissue (6 out of 28 vs. 0 out of 17; Fisher test: p = 0.0463). The worst-case survival rate with the removal of the arthroplasty for any cause and/or loss to follow-up was 80.0% (95% CI: 44.9–100) at 74.7 months (n = 1 due to low-grade infection). Conclusions: En-bloc resection of metastases and reconstruction by modular arthroplasty is reliable even in patients with very limited life expectancy. Local complications due to tumor growth or instability after intralesional surgery could be managed successfully but recurrent dislocation as the most frequent complication has to be taken into account. The simple bonding of remaining soft tissue around the prosthesis without the use of an attachment tube may reduce the dislocation rate and reoperation risk.
Frequently occurring damage, as well as elevated blood metal ion levels, are reported in relation to a tumor and revision system for total knee arthroplasty (TKA), which applies a rotating hinge coupling with a metal-on-metal (MoM) articulation. As the patient collective for this specific system is small, there is no data on wear generated from the couplings. In this study, wear volume and influencing parameters were investigated at 44 retrieved TKAs with MoM couplings. A scoring system rating frequently occurring abrasive wear between 0 (no wear) and 3 (distinct wear) was established. The wear score was correlated to time in vivo, bone resection length, patient weight and polyethylene inlay damage. Volumetric wear was estimated applying coordinate measurements. An elevated wear score of two or higher was found in 43% of cases. The mean wear rate accounted to 7.8 mm3/year. The main influencing coefficient for the extent of wear is time in vivo. We found a tendency for higher wear scores with higher inlay degradation scores. Patient weight and bone resection length did not impact coupling wear. Assessment of wear damage by a semi-quantitative scoring system has proven to be a reliable option for non-destructive coupling evaluation. The generated wear volume is high.
Adequate primary stability of the acetabular revision construct is necessary for long-term implant survival. The difference in primary stability between tantalum and titanium components is unclear. Six composite hemipelvises with an acetabular defect were implanted with a tantalum augment and cup, using cement fixation between cup and augment. Relative motion was measured at cup/bone, cup/augment and bone/augment interfaces at three load levels; the results were compared to the relative motion measured at the same interfaces of a titanium cup/augment construct of identical dimensions, also implanted into composite bone. The implants showed little relative motion at all load levels between the augment and cup. At the bone/augment and bone/cup interfaces the titanium implants showed less relative motion than tantalum at 30% load (p < 0.001), but more relative motion at 50% (p = n.s.) and 100% (p < 0001) load. The load did not have a significant effect at the augment/cup interface (p = 0.086); it did have a significant effect on relative motion of both implant materials at bone/cup and bone/augment interfaces (p < 0.001). All interfaces of both constructs displayed relative motion that should permit osseointegration. Tantalum, however, may provide a greater degree of primary stability at higher loads than titanium. The clinical implication is yet to be seen
AbstractFemoral component loosening is a rare but severe complication in total knee arthroplasty. Former studies have repeatedly demonstrated radiolucent lines behind the ventral and dorsal anchoring shields of the femoral components, which has led us to investigate this matter further. Therefore, three different cementing techniques were tested in a group of nine Sawbone samples each. These differed in the amount of cement applied on the femoral component as well as in the pressure application. Computed tomography was performed to evaluate and classify the cement penetration into the bone adjacent to the prosthesis according to the zones defined by the Knee Society scoring system. The results show significantly deeper cement penetration in all zones when a pressurizer is used. In the other two groups, no significant difference in the dorsal bevel cement penetration was noted. Additionally, no difference in ventral and dorsal cement penetrations (Zones 1 and 4) was delineated. In contrast, there was a significant difference in both the ventral bevel (Zone 2) as well as the distal anchoring surface (Zones 5–7). The use of a pressurizer results in greater cement penetration into all anchoring areas. Completely covering the component back surface results in a significantly higher penetration, which is mainly due to differences in volume. These data show significantly improved cementation results when using a pressurizer. Whether this improves the biomechanical properties and ultimately the revision rate requires further investigation.
Primary total hip arthroplasty (THA) is one of the most successful surgical procedures. Considering the demographic change the use of new ultra-short femoral implants has gained importance especially when treating young patients. Main features are bone conservation, metaphyseal anchoring and thus reducing stress shielding by proximal load transferring. The objective of this study is to give an overview over the subject of femoral neck prostheses. A systematic review was conducted. A total number of 27 publications were taken into this systematic review. Over all, just a few follow-up, biomechanical and radiostereometric studies have been conducted in the past. Still no long-term results (>10 years of follow-up) are available. The available mid-term results indicate unsatisfactory survival rates. Aseptic loosening was the most common reason for revision. Valgus angle and good bone mineral density were considered to be crucial for primary stability of femoral neck prostheses. Register data report a very low percentage of femoral neck prostheses in THA with even more diminishing implantation rates. To conclude, further studies are necessary in order to provide evidence-based recommendations. Currently, due to the inhomogeneous and poor data a reasonable and legitimate recommendation cannot be given.
Background: The twin peg femoral component was introduced for the cemented Oxford unicondylar knee to increase implant stability. The aim of this experimental study was to investigate the influence of the twin peg design on femoral interface temperature and maximum load to failure in comparison to the single peg design. Methods: In this experimental study medial Oxford unicompartmental knee arthroplasty was performed in 12 pairs of fresh-frozen human knees. A cemented femoral single peg component was implanted on the one side (group A) and a cemented twin peg component on the other side (group B). Cement interface temperature was continuously monitored during the procedure. Maximum tensile forces of the femoral components were measured by pull-out tests. Findings: Maximum femoral interface temperatures did not reach critical values for heat necrosis of the bone in group A (mean 28.4, SD 1.2 degrees C) or group B (mean 27.6, SD 0.5 degrees C). The maximum load to failure was significantly higher in the twin peg group (mean 3628.41, SD 650.92 N) compared to the single peg group (mean 2979, SD 781 N) (P = 0.016). Interpretation: Our experiments showed higher load to failure for the twin peg design compared to the single peg design without raising the risk of heat necrosis at the interfacial bone. The twin peg component offers a save alternative to the single peg component in a cadaveric setting.
Background: In uncemented total hip arthroplasty (THA), low bone mineral density (BMD) is associated with aseptic loosening. BMD is usually assessed via dual-energy X-ray absorptiometry (DXA) or quantitative computed tomography, which takes time and exposes patients to radiation. Due to its low risk profile, intraoperative measurement of the trabecular stability might be a useful alternative to DXA. Methods: In 24 human femora, BMD was analysed using DXA at the femoral necks and the knees. Performing the standard Oxford Unicompartmental Knee Arthroplasty (OUKA) implantation procedure, a wingblade (DensiProbe) coupled to a torque probe was used to evaluate the trabecular peak torque. The standard procedure was modified: before the completion of the central peg drill hole, the DensiProbe was inserted into the pre-drilled hole and then turned until a loss of resistance was achieved. The obtained data was then correlated with BMD at the femoral neck as well as the knee. Results: In all tested regions, a higher peak torque was observed in correlation with a higher BMD. Conclusions: As demonstrated, the DensiProbe can be a helpful tool to assess the bone quality intraoperatively in OUKA. It can be a valuable decision guidance when faced with choosing between a cemented and a cementless implant. Due to the fact that the central peg hole of the OUKA can be used for the procedure, no additional risk for the patient exists, while the additional work for the surgeon is minimal. (C) 2018 Elsevier B.V. All rights reserved.
Purpose of the present cohort study was the determination of lower body function and rotation in patients with symptomatic component mal-rotation after total knee arthroplasty using instrumented 3D gait analysis.
Background: Revision total hip arthroplasty frequently faces challenges associated with deficient bone stock. Porous metal implants were developed to meet the challenge, but require rapid osseointegration for ultimate success. This study aims to assess relative motion as an indicator for primary stability and osseointegration of two different titanium cups each combined with a titanium augment. Methods: In 14 cadaver hemipelvises, 2 types of titanium acetabular cups, a traditional sintered-bead cup (POROCOAT Acetabular Cup [PAC]) and a newer porous-coated cup (GRIPTION Acetabular Cup [GAC]) each associated with a porous augment, were subjected to 3-dimensional varying loads, replicating 30% of loads experienced during normal gait. Relative motion was measured at the cup/bone, augment/bone, and cup/augment interfaces. Results: Only at the cup/bone interface was there a statistically significant difference in relative motion between the traditional PAC and the newer GAC, with PAC showing less relative motion (P .0037). Bone mineral density (BMD) had a significant effect on relative motion (P=.0019) at the cup/bone interface of both cup types, with low BMD specimens showing more relative motion. Conclusion: Both cup types combined with augments displayed minimal relative motion that was within the accepted range thought to allow osseointegration, although the traditional surface proved superior to the newer surface. This difference was more pronounced at low BMD, with the well-established PAC cup displaying less relative motion than the more porous GAC cup, consistent with better osseointegration than the more porous cup. This suggests that the more porous implant may be less advantageous than traditional PAC cups, particularly in cases with poorer bone stock. (C) 2018 Elsevier Inc. All rights reserved.
OBJECTIVES:In order to address acetabular defects, porous metal revision acetabular components and augments have been developed, which require fixation to each other. The fixation technique that results in the smallest relative movement between the components, as well as its influence on the primary stability with the host bone, have not previously been determined. METHODS:A total of 18 composite hemipelvises with a Paprosky IIB defect were implanted using a porous titanium 56 mm multihole acetabular component and 1 cm augment. Each acetabular component and augment was affixed to the bone using two screws, while the method of fixation between the acetabular component and augment varied for the three groups of six hemipelvises: group S, screw fixation only; group SC, screw plus cement fixation; group C, cement fixation only. The implanted hemipelvises were cyclically loaded to three different loading maxima (0.5 kN, 0.9 kN, and 1.8 kN). RESULTS:Screw fixation alone resulted in up to three times more movement (p = 0.006), especially when load was increased to 100% (p < 0.001), than with the other two fixation methods (C and SC). No significant difference was noted when a screw was added to the cement fixation. Increased load resulted in increased relative movement between the interfaces in all fixation methods (p < 0.001). CONCLUSION:Cement fixation between a porous titanium acetabular component and augment is associated with less relative movement than screw fixation alone for all implant interfaces, particularly with increasing loads. Adding a screw to the cement fixation did not offer any significant advantage. These results also show that the stability of the tested acetabular component/augment interface affects the stability of the construct that is affixed to the bone.Cite this article: N. A. Beckmann, R. G. Bitsch, M. Gondan, M. Schonhoff, S. Jaeger. Comparison of the stability of three fixation techniques between porous metal acetabular components and augments. Bone Joint Res 2018;7:282-288. DOI: 10.1302/2046-3758.74.BJR-2017-0198.R1.
As there are many reports describing avascular reactions to metal debris (ARMD) after Metal-on-Metal Hip Arthroplasty (MoMHA), the use of MoMHA, especially hip resurfacing, is decreasing worldwide. In cases of ARMD or a rise of metal ion blood levels, revision is commended even in pain free patients with a well integrated implant. The revision of a well integrated implant will cause bone loss. As most of the patients with a hip resurfacing are young and a good bone stock is desirable for further revision surgeries, the purpose of this study was to evaluate the stability of a cemented polyethylene cup in a metal hip resurfacing cup. Two different hip resurfacing systems were investigated in this study (ASR™, DePuy Orthopaedics, Leatherhead, UK; Cormet™, Corin Group, Cirencester, UK).Six different groups were formed according to the treatment and preparation of the cement-cup-interface (table 1). Before instilling cement in groups 1, 3, 5 the surface, which was contaminated with blood, was cleaned just usin...
IntroductionThe frequency of revision hip arthroplasty is increasing with the increasing life expectancy and number of individuals treated with joint replacement. Newer porous implants have been introduced which may provide better treatment options for revision arthroplasty. These may require cementation to other prosthesis components and occasionally to bone, however, there is currently no information on how these porous implants interface with cement.Materials and MethodsCylindrical bone (control group) and porous metal probes with a diameter and height of 10mm were created and subsequently cemented in a standardized setting. These were placed under tensile and torsional loading scenarios.In this experimental study, 10 human femoral heads were used to create 20 cylindrical probes with a diameter and height of 10mm. One side was tapered to 6mm for cementation and interface evaluation. A further set of 20 probes of a porous metal implant (Trabecular Metal®) was created with the same geometry. After the pr...